Sweat is a promising non-invasive biofluid that can provide valuable insights into the physiological state of the human body. However, a major obstacle to analyzing sweat in real-time is the fabrication of simple, fast-acting, accurate, and low-cost sensing constructs. To address this challenge, we introduced easily-prepared wearable hydrogel sensors that can be placed on the skin and used colorimetric techniques to assess sweat analytes without invasive procedures. Two typical sweat sensors, chloride ion (Cl) responsive patches for cystic fibrosis (CF) analysis and glucose response patches for diabetic monitoring, were demonstrated for real sample analysis. The Cl colorimetric sensor, with a detection limit down to 100 μM, shows a good linear response from 1.56 mM to 200 mM Cl, and the glucose colorimetric sensor, with a detection limit down to 1 μM, exhibits an adequate linear response from 10 μM to 1 mM glucose. These colorimetric hydrogel sensors are also incorporated into a medical dressing to create wearable sensor devices for real-time sweat analysis. The acquired readings closely match the results obtained from the benchmark analyzing instrument, with a small deviation of less than 10%. Therefore, our simple colorimetric hydrogel sensing patches hold great potential to advance real-time sweat testing and contribute to the transitional development of wearable medical devices.