Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: Quantitative human chorionic gonadotropin (hCG) measurements are used to manage women classified with a pregnancy of unknown location (PUL). Two point of care testing (POCT) devices that quantify hCG are commercially available. We verified the i-STAT 1 (Abbott) and the AQT 90 FLEX (Radiometer) prior to use in PUL triage.
Methods: Tests for precision, external quality assurance (EQA), correlation, hook effect and recovery were undertaken alongside a POCT usability assessment during this prospective multi-center verification.
Results: Coefficients of variation ranged between 4.0 and 5.1 % for the three i-STAT 1 internal quality control (IQC) solutions and between 6.8 and 7.3 % for the two AQT IQC solutions. Symmetric differences in POCT EQA results when compared with laboratory and EQA stock values ranged between 3.2 and 24.5 % for the i-STAT 1 and between 3.3 and 36.9 % for the AQT. Correlation coefficients (i-STAT 1: 0.96, AQT: 0.99) and goodness of fit curves (i-STAT 1: 0.92, AQT: 0.99) were excellent when using suitable whole blood samples. An hCG hook effect was noted with the i-STAT 1 between 572,194 and 799,089 IU/L, lower than the hook effect noted with the AQT, which was between 799,089 and 1,619,309 IU/L. When hematocrit concentration was considered in sample types validated for use with each device, hCG recovery was 108 % with the i-STAT 1 and 98 % with the AQT. The i-STAT 1 scored lower on usability overall (90/130) than the AQT (121/130, p<0.001, Mann-Whitney).
Conclusions: Both hCG POCT devices were verified for use in clinical practice. Practical factors must also be considered when choosing which device to use in each unit.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/cclm-2023-0703 | DOI Listing |