Purpose: Recent reports that pulse oximeters may overestimate oxygen saturation in individuals with darker skin pigmentation have prompted concerns from regulatory authorities regarding racial bias. We investigated the performance of TruSignal SpO2 sensors (GE Healthcare, Helsinki, Finland) in adults with varying skin pigmentation.

Methods: A retrospective study was conducted using a set of pooled assessments of SpO2/SaO2 measurements from nine studies to assess bias, accuracy (A), and precision of TruSignal sensors in healthy adults under induced hypoxia. Subgroup analyses were performed based on oxygen saturation levels (band 1, ≥ 70 and ≤ 80%; band 2, > 80 and ≤ 90%; band 3, > 90 and ≤ 100%).

Results: Of the 10,800 data points from 131 individuals, 8,202 (75.9%) and 2,598 (24.1%) were assigned to the light and dark pigment groups, respectively. Bias was 0.14% overall and less than 1% across oxygenation bands. The difference in bias between dark and light pigment groups was statistically significant at the low oxygenation band with SpO2 ≥ 70 and ≤ 80% (+ 0.58% and + 0.30% respectively; p = 0.0035). Throughout the saturation range, A was 1.64% in the light and 1.71% in the dark pigment group, within device specifications and regulatory requirements. Oxygenation was the dominating factor in stepwise ANOVA modeling. The mixed model also showed that bias was strongly affected by the oxygenation range.

Conclusion: TruSignal sensors demonstrated higher bias at lower oxygen saturation, with less than 0.5% difference between pigment groups. These findings raise new questions, such as ways to improve pulse oximetry measurements during challenging clinical conditions, including low perfusion.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10879215PMC
http://dx.doi.org/10.1007/s10877-023-01095-1DOI Listing

Publication Analysis

Top Keywords

oxygen saturation
12
pigment groups
12
pulse oximetry
8
trusignal sensors
8
dark pigment
8
bias
7
assessment skin
4
skin pigmentation-related
4
pigmentation-related bias
4
bias pulse
4

Similar Publications

Objective: There is a connection between obstructive sleep apnea (OSA) and coronary microvascular dysfunction (CMD), but the underlying mechanisms remain unclear. This study aims to evaluate the correlation between OSA-related nocturnal hypoxemia parameters and CMD.

Methods: This is an observational, single-center study that included patients who underwent polysomnography and coronary angiography during hospitalization.

View Article and Find Full Text PDF

Context-aware analysis enhances autoscoring accuracy of home sleep apnea testing.

J Clin Sleep Med

January 2025

Natural Interaction Lab, Thom Building, Department of Engineering, University of Oxford, Oxford, United Kingdom.

Study Objectives: Home sleep apnea testing based on peripheral arterial tonometry (P-HSAT) is increasingly being deployed because of its ability to test for multiple nights. However, P-HSATs do not have access to modalities such as airflow and cortical arousals and instead rely on alternative sources of information to detect respiratory events. This results in an a-priori performance disadvantage.

View Article and Find Full Text PDF

Right ventricular myocardial work: proof-of-concept for the assessment of pressure-strain loops of patients with pre-capillary pulmonary hypertension.

Cardiovasc Ultrasound

January 2025

Cardiology Department, Hospital de Santa Marta, Unidade Local de Saúde São José, Centro Clínico Académico de Lisboa, Rua de Santa Marta N.º 50, Lisbon, 1169-024, Portugal.

Background: Right ventricular myocardial work (RVMW) assessed by transthoracic echocardiography allows to study the right ventricular (RV) function using RV pressure-strain loops. The assessment of these novel indexes of RVMW has not yet been exten sively studied, namely in pre-capillary pulmonary hypertension (PH) population.

Objectives: to evaluate the relationship between RVMW and invasive indices of right heart catheterization (RHC) in a cohort of patients with group I and group IV PH and to compare with a control group without PH.

View Article and Find Full Text PDF

Cerebral oxygen saturation in cirrhotic patients with gastro-intestinal bleeding, a near infrared spectroscopy study.

Metab Brain Dis

January 2025

Brain Liver Pitié-Salpêtrière (BLIPS) Study Group, Centre de recherche Saint- Antoine, Maladies métaboliques, biliaires et fibro-inflammatoire du foie, Institute of Cardiometabolism and Nutrition (ICAN), INSERM UMR_S 938, Paris, France.

Near Infrared Spectroscopy (NIRS) is a non-invasive optical technique allowing a continuous measurement of brain's hemoglobin (Hb) saturation in oxygen (rSO2). It is a marker of cerebral insult and rSO2 < 50% is associated with increased neurological impairment. Cirrhotic patients with gastrointestinal bleeding (GIB) often develop hepatic encephalopathy (HE).

View Article and Find Full Text PDF

Incremental shuttle walking test for calf muscle oxygenation assessment in peripheral arterial disease: a cross-sectional study.

Sci Rep

January 2025

Graduate Program in Rehabilitation Sciences, Universidade Federal de Minas Gerais, Avenida Presidente Antônio Carlos, 6627 - Pampulha, Belo Horizonte, CEP 31270-901, MG, Brazil.

People with peripheral arterial disease (PAD) and intermittent claudication (IC) experience impaired walking due to an imbalance between muscle oxygen supply and demand during exercise. Studies with near-infrared spectroscopy (NIRS) during treadmill tests reveal notable tissue deoxygenation with slow recovery. This cross-sectional study aimed to compare behavior of calf muscle oxygenation during the incremental shuttle walking test (ISWT) with a continuous treadmill test (3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!