Biomechanical causes for failure of the Physiomesh/Securestrap system.

Sci Rep

Department of General Surgery and Hernia Center, Swissmed Hospital, Wileńska 44, 80-215, Gdańsk, Poland.

Published: October 2023

This study investigates the mechanical behavior of the Physiomesh/Securestrap system, a hernia repair system used for IPOM procedures associated with high failure rates. The study involved conducting mechanical experiments and numerical simulations to investigate the mechanical behavior of the Physiomesh/Securestrap system under pressure load. Uniaxial tension tests were conducted to determine the elasticity modulus of the Physiomesh in various directions and the strength of the mesh-tissue-staple junction. Ex-vivo experiments on porcine abdominal wall models were performed to observe the system's behavior under simulated intra-abdominal pressure load. Numerical simulations using finite element analysis were employed to support the experimental findings. The results reveal nonlinearity, anisotropy, and non-homogeneity in the mechanical properties of the Physiomesh, with stress concentration observed in the polydioxanone (PDO) stripe. The mesh-tissue junction exhibited inadequate fixation strength, leading to staple pull-out or breakage. The ex-vivo models demonstrated failure under higher pressure loads. Numerical simulations supported these findings, revealing the reaction forces exceeding the experimentally determined strength of the mesh-tissue-staple junction. The implications of this study extend beyond the specific case of the Physiomesh/Securestrap system, providing insights into the mechanics of implant-tissue systems. By considering biomechanical factors, researchers and clinicians can make informed decisions to develop improved implants that mimic the mechanics of a healthy abdominal wall. This knowledge can contribute to better surgical outcomes and reduce complications in abdominal hernia repair and to avoid similar failures in future.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10579252PMC
http://dx.doi.org/10.1038/s41598-023-44940-8DOI Listing

Publication Analysis

Top Keywords

physiomesh/securestrap system
16
numerical simulations
12
mechanical behavior
8
behavior physiomesh/securestrap
8
hernia repair
8
pressure load
8
strength mesh-tissue-staple
8
mesh-tissue-staple junction
8
abdominal wall
8
system
5

Similar Publications

Biomechanical causes for failure of the Physiomesh/Securestrap system.

Sci Rep

October 2023

Department of General Surgery and Hernia Center, Swissmed Hospital, Wileńska 44, 80-215, Gdańsk, Poland.

This study investigates the mechanical behavior of the Physiomesh/Securestrap system, a hernia repair system used for IPOM procedures associated with high failure rates. The study involved conducting mechanical experiments and numerical simulations to investigate the mechanical behavior of the Physiomesh/Securestrap system under pressure load. Uniaxial tension tests were conducted to determine the elasticity modulus of the Physiomesh in various directions and the strength of the mesh-tissue-staple junction.

View Article and Find Full Text PDF

Background: Patients' need to improve outcomes and to reduce the number of complications triggers the development of new materials and surgery concepts. Currently, there are many implants and fixation systems dedicated for intraperitoneal onlay mesh procedure. The aim of this study was to compare two different mesh/fixation system concepts (PH: Physiomesh/Securestrap and VS: Ventralight ST/SorbaFix) for laparoscopic ventral hernia repair with respect to pain.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!