Background: Hitherto, lumen less leads (LLLs) were routinely utilized for conduction system pacing (CSP). We report the largest experience using stylet-driven leads (SDLs) with a deflectable mapping catheter for CSP.

Methods: Patients were prospectively and sequentially enrolled for CSP with SDL between June, 2021 and November, 2022 to (i) a novel deflectable mapping catheter (AgilisHisPro, Abbott) (Group A) or (ii) a fixed curve sheath (Selectra3D, Biotronik) (Group B) in a 1:1 non-randomized fashion. The primary aim was to evaluate safety, feasibility, and efficacy of the CSP using SDL and deflectable mapping catheter (Group A) while reporting procedural success and intermediate-term follow-up.

Results: Seventy-nine patients (59.4%M, mean age 67.2+/-10.6 years) were allocated to either (i) Group A (n = 40) or (ii) Group B (n = 39). In Group A (n = 40, 50% M, mean age 67.2+/-9.5 years, follow-up 210.7 + 25.1days), His bundle pacing (HBP) was the default strategy with left bundle branch area pacing (LBBaP) for bailout. Procedural success with HBP was feasible in 17/40 (42.5%) patients with remaining 23/40 (57.5%) needing LBBaP bailout. After initial learning curve, a manual septal curve was introduced to successfully aid LBBaP in 6/23 (26.1%) cases. Procedural and follow-up parameters did not differ significantly in HBP vs. LBBaP. Head-to-head comparison was not performed between the groups owing to different default protocols (HBP-Group A, Discretionary-Group B).

Conclusions: Use of SDL with single-curve deflectable mapping catheter was safe, feasible and yielded moderate procedural success with HBP and frequently needed a LBBaP bailout strategy. In approximately one-fourth of the latter, an out-of-plane manual septal curve was needed to optimize LBBaP.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10840-023-01637-2DOI Listing

Publication Analysis

Top Keywords

deflectable mapping
20
mapping catheter
20
procedural success
12
lbbap bailout
12
conduction system
8
system pacing
8
csp sdl
8
group group
8
success hbp
8
manual septal
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!