Skin surface contamination by alpha-emitters is in itself not hazardous, but it would cause significant internal exposure in the case of injured skin as well as misjudgment in direct in vivo measurements (e.g. lung counting). The present study determined the source efficiency of alpha-emitters (241Am) applied to swine skin samples by analysing the observed alpha-particle energy spectra using advanced alpha-spectrometric simulation. Based on our results, the source efficiency was determined to be 0.365 (alpha-particle s-1 per Bq) on average (c.f. 0.5 in the case of no self-absorption in the source). The decrease in source efficiency would be attributed primarily to the radionuclide entering hair follicles or deep wrinkles. The degradation of the measured spectra from the skin samples indicates the penetration of some radionuclides into the upper layers of the stratum corneum. Although this study was limited to results obtained from swine skin samples, it suggests that irregularities in the skin surface may affect direct alpha measurements.