Photon-Counting Detector CT for Femoral Stent Imaging in an Extracorporeally Perfused Human Cadaveric Model.

Invest Radiol

From the Department of Diagnostic and Interventional Radiology, University Hospital of Würzburg, Würzburg, Germany (V.H., P.G., H.H., A.M.A., T.A.B., J.-P.G., B.M.W.P.); Institute of Anatomy and Cell Biology, Julius-Maximilians University Würzburg, Würzburg, Germany (F.K., S.E.); and Department of General, Visceral, Transplant, Vascular, and Pediatric Surgery, Center of Operative Medicine, University Hospital of Würzburg, Würzburg, Germany (D.P., S.L.).

Published: April 2024

Background And Aims: This study aims to compare the performance of first-generation dual-source photon-counting detector computed tomography (PCD-CT) to third-generation dual-source energy-integrating detector (EID-CT) regarding stent imaging in the femoral arterial runoff.

Methods: Continuous extracorporeal perfusion was established in 1 human cadaver using an inguinal and infragenicular access and peristaltic pump. Seven peripheral stents were implanted into both superior femoral arteries by means of percutaneous angioplasty. Radiation dose-equivalent CT angiographies (high-/medium-/low-dose: 10/5/3 mGy) with constant tube voltage of 120 kVp, matching iterative reconstruction algorithm levels, and convolution kernels were used both with PCD-CT and EID-CT. In-stent lumen visibility, luminal and in-stent attenuation as well as contrast-to-noise ratio (CNR) were assessed via region of interest and diameter measurements. Results were compared using analyses of variance and regression analyses.

Results: Maximum in-stent lumen visibility achieved with PCD-CT was 94.48% ± 2.62%. The PCD-CT protocol with the lowest lumen visibility (BV40: 78.93% ± 4.67%) performed equal to the EID-CT protocol with the best lumen visibility (BV59: 79.49% ± 2.64%, P > 0.999). Photon-counting detector CT yielded superior CNR compared with EID-CT regardless of kernel and dose level ( P < 0.001). Maximum CNR was 48.8 ± 17.4 in PCD-CT versus 31.28 ± 5.7 in EID-CT (both BV40, high-dose). The theoretical dose reduction potential of PCD-CT over EID-CT was established at 88% (BV40), 83% (BV48/49), and 73% (BV59/60), respectively. In-stent attenuation was not significantly different from luminal attenuation outside stents in any protocol.

Conclusions: With superior lumen visibility and CNR, PCD-CT allowed for noticeable dose reduction over EID-CT while maintaining image quality in a continuously perfused human cadaveric model.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0000000000001019DOI Listing

Publication Analysis

Top Keywords

lumen visibility
20
photon-counting detector
12
stent imaging
8
perfused human
8
human cadaveric
8
cadaveric model
8
pcd-ct eid-ct
8
in-stent lumen
8
in-stent attenuation
8
dose reduction
8

Similar Publications

Objective: To investigate the protective effect of lycopene on lung oxidative damage induced by atmospheric fine particulate matter(PM_(2.5)) in rats.

Methods: Sixty 7-week-old male Sprague-Dawley rats were randomly divided into six groups: normal control group, PM_(2.

View Article and Find Full Text PDF

Objectives: Pressurized IntraThoracic Aerosol Chemotherapy (PITAC) has been suggested as a new therapy for patients with malignant pleural effusion (MPE) and/or pleural metastasis (PLM). The patients have a poor prognosis with a median survival of 3 to 12 months. We present feasibility, patient safety, and cytological/histological response assessment in PITAC-treated patients with MPE and/or PLM.

View Article and Find Full Text PDF

Purpose: This paper outlines the commissioning of the Varian (VMS, Varian Medical Systems, Palo Alto, CA) Universal Interstitial Cylinder (UIC) applicator set for Ir-192 HDR brachytherapy. The UIC was commissioned for use with CT and MRI and a custom phantom was designed to avoid the introduction of water-like materials into the needle guide tracks. Various marker strands were investigated to determine which allowed the most accurate reconstruction of source positions.

View Article and Find Full Text PDF

This study evaluated the growth performance, immune response, hepatopancreatic health, and disease resistance in fed diets supplemented with fermented pomegranate peel polyphenols (FPPP) for 45 days. Five diets were formulated to contain various levels of FPPP: FP0 (no FPPP), FPPP inclusion at 0.015% (FP1), 0.

View Article and Find Full Text PDF
Article Synopsis
  • * Fiber-optic laryngoscopy showed a highly mobile lesion affecting the vocal cords and a very small tracheal lumen visible only during expiration.
  • * The patient was successfully intubated awake using a specialized tube over a pediatric bronchoscope, showcasing the importance of thorough preoperative assessment and adherence to difficult airway protocols to prevent complications.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!