Effect of scan powder and scanning technology on measured deviations of complete-arch implant supported frameworks digitized with industrial and intraoral scanners.

J Dent

Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland; Department of Restorative, Preventive and Pediatric Dentistry, School of Dental Medicine, University of Bern, Bern, Switzerland; Division of Restorative and Prosthetic Dentistry, The Ohio State University College of Dentistry, Columbus, Ohio, USA.

Published: November 2023

Objectives: To evaluate the suitability of intraoral scanners (IOSs) to analyze the fabrication trueness of titanium complete-arch implant-supported frameworks by comparing with an industrial-grade scanner and investigate how anti-reflective scan powder affects measured deviations.

Methods: Ten titanium complete-arch implant-supported frameworks were milled from a reference standard tessellation language (STL) file. An industrial-grade blue light scanner (ATOS Core 80 (AT)) and three IOSs (Primescan (PS), TRIOS T3 (T3), and TRIOS T4 (T4)) with (PS-P, T3-P, and T4-P) or without (PS, T3, and T4) anti-reflective scan powder application were used to generate test STL (TSTL) files of the frameworks. Reference STL and TSTLs were imported into a metrology-grade analysis software (Geomagic Control X) and whole surface root mean square (RMS) values were calculated. Another software (Medit Link v 2.4.4) was used to virtually isolate marginal surfaces of all STL files and marginal RMS values were calculated by using the same metrology-grade analysis software. A linear mixed effects model was used to compare the transformed deviations of the scans performed by using each IOS (with or without powder) with the deviations of those performed by using the reference AT scanner within each surface, where a Box-Cox type transformation was used for variance stability. Bonferroni corrected post-hoc tests were used to compare conditions within each IOS (α=0.05).

Results: All IOSs had significantly higher whole surface and marginal RMS values than AT, regardless of the condition (P≤.002). However, scan powder application did not affect the whole surface and marginal RMS values in scans of tested IOSs (P≥.054).

Conclusion: Measured whole surface and marginal deviations in all IOS scans performed with or without the use of scan powder were higher than those in AT scans. The application of anti-reflective scan powder did not affect the deviations in scans of tested IOSs.

Clinical Significance: Even though deviations measured in the scans of tested scanners were significantly different than those in the reference scanner, the maximum raw mean difference was 37.33 µm and the maximum raw confidence interval value of estimated differences was 47.88 µm, which can be considered clinically small taking into account the size of the frameworks tested. Therefore, tested intraoral scanners may be feasible to scan prostheses similar to or smaller than tested frameworks for fabrication trueness analysis, which may facilitate potential clinical adjustments.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jdent.2023.104736DOI Listing

Publication Analysis

Top Keywords

scan powder
24
rms values
16
intraoral scanners
12
anti-reflective scan
12
marginal rms
12
surface marginal
12
scans tested
12
fabrication trueness
8
titanium complete-arch
8
complete-arch implant-supported
8

Similar Publications

Facile phase selective synthesis of copper antimony sulphide (CAS) nanostructures is important because of their tunable photoconductive and electrochemical properties. In this study, off-stoichiometric famatinite phase CAS (CAS) quasi-spherical and quasi-hexagonal colloidal nanostructures (including nanosheets) of sizes, 2.4-18.

View Article and Find Full Text PDF

In the realm of materials science and engineering, the pursuit of advanced materials with tailored properties has been a driving goal behind technological progress. Scientific interest in laser powder bed fusion (L-PBF) fabricated NiTi alloy has in recent times seen an upsurge of activity. In this study, we investigate the impact of varying volume energy density (VED) during L-PBF on the microstructure and corrosion behaviour of NiTi alloys in both scan (XY) and built (XZ) planes.

View Article and Find Full Text PDF

Graphene aerogels (GAs) with engineered architectures are a promising material for applications ranging from filtration to energy storage/conversion. However, current preparation approaches involve the combination of multiple intrinsically-different methodologies to achieve graphene-synthesis and architecture-engineering, complicating the entire procedure. Here, a novel approach to prepare GAs with engineered architectures based on the laser-upcycling of protein biowaste, hemoglobin, is introduced.

View Article and Find Full Text PDF

This study examined scan speed, hatch distance, and scan rotation angle parameters to determine the effects of the powder bed fusion process on the tensile strength of AlSi10Mg material. The Taguchi L9 experimental design was applied to evaluate the effects of the parameters systematically. The experimental results revealed the importance of the parameters affecting the tensile strength of the AlSi10Mg material.

View Article and Find Full Text PDF

In this study, DL-phenylalanine modified with a multiwall carbon nanotube paste electrode is used as advanced electrochemical sensor for analysing of 0.1 mM caffeic acid (CFA) with simultaneous detection of riboflavin (RFN). The developed sensors include electrochemically polymerized DL-phenylalanine (DL-PA) modified multiwall carbon nanotube paste electrode [DL-PAMMCNTPE] and bare multiwall carbon nanotube paste electrode [BMCNTPE].

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!