Escape of surgical smoke particles, comparing conventional and valveless trocar systems.

Surg Endosc

Department of Biomechanical Engineering, Faculty of Mechanical Engineering, Delft University of Technology, Mekelweg 2, 2628 CD, Delft, The Netherlands.

Published: November 2023

Background: During minimal access surgery, surgical smoke is produced which can potentially be inhaled by the surgical team, leading to several health risks. This smoke can escape from the abdominal cavity into the operating room due to trocar leakage. The trocars and insufflator that are used during surgery influence gas leakage. Therefore, this study compares particle escape from a valveless (Conmed AirSeal iFS), and a conventional (Karl Storz Endoflator) system.

Materials And Methods: Using an in vitro model, a conventional and a valveless trocar system were compared. A protocol that simulated various surgical phases was defined to assess the surgical conditions and particle leakage. Insufflation pressures and instrument diameters were varied as these are known to affect gas leakage.

Results: The conventional trocar leaked during two distinct phases. Removal of the obturator caused a sudden release of particles. During instrument insertion, an average of 211 (IQR 111) particles per second escaped when using the 5 mm diameter instrument. With the 10 mm instrument, 50 (IQR 13) particles per second were measured. With the conventional trocar, a higher abdominal pressure increased particle leakage. The valveless trocar demonstrated a continuously high particle release during all phases. After the obturator was removed, particle escape increased sharply. Particle escape decreased to 1276 (IQR 580) particles per second for the 5 mm instrument insertion, and 1084 (IQR 630) particles per second for 10 mm instrument insertion. With the valveless trocar system, a higher insufflation pressure lowered particle escape.

Conclusions: This study shows that a valveless trocar system releases more particles into the operating room environment than a conventional trocar. During instrument insertion, the leakage through the valveless system is 6 to 20 times higher than the conventional system. With a valveless trocar, leakage decreases with increasing pressure. With both trocar types leakage depends on instrument diameter.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10615925PMC
http://dx.doi.org/10.1007/s00464-023-10444-9DOI Listing

Publication Analysis

Top Keywords

valveless trocar
24
instrument insertion
16
particles second
16
particle escape
12
trocar system
12
conventional trocar
12
trocar
11
surgical smoke
8
valveless
8
conventional valveless
8

Similar Publications

Background: Valveless trocars like AirSeal system are maintain a stable pneumoperitoneum and reduce instrument friction.

Case Presentation: A 65-year-old man's robotic radical cystectomy was complicated by a missing needle while using AirSeal system. The needle was detected via backward inspection using the endo camera inserted through the trocar, tip at its most distal end let the visualization of the needle within the air channels and confirmed with a trocar X-ray.

View Article and Find Full Text PDF

A systematic review and meta-analysis were performed to investigate the efficacy of the AirSeal Valveless Trocar Needle Insufflation System in robot-assisted partial nephrectomy (RAPN). The study compared the differences in perioperative outcomes between the AirSeal insufflation group (AIS) and the conventional insufflation group (CIS). A systematic search of databases such as PubMed, Embase, Cochrane library, and Web of science was performed to identify studies reporting perioperative outcomes between the AirSeal insufflation group (AIS) and the conventional insufflation group (CIS) in RAPN.

View Article and Find Full Text PDF

Purpose: AirSeal is a valve-less trocar insufflation system which is widely used in robotic urologic surgeries. More evidence is needed concerning the application and cost of AirSeal in retroperitoneal robot-assisted laparoscopic partial nephrectomy.

Methods: We conducted a randomized controlled trial enrolling 62 patients who underwent retroperitoneal robot-assisted laparoscopic partial nephrectomy from February 2022 to February 2023 in the Peking Union Medical College Hospital.

View Article and Find Full Text PDF

Background: Robotic mitral repair is generally performed with four intercostal trocars and a minithoracotomy. We describe our technique and results with a totally-thoracoscopic closed chest approach using a 12 mm valveless trocar as "working port", without a minithoracotomy. We compared our results with this technique with a control group of robotic mitral repairs performed earlier with a minithoracotomy.

View Article and Find Full Text PDF

Conventional operative insufflation uses a one-way trocar to handle instruments while maintaining pneumoperitoneum. In 2007, the AirSeal valveless trocar insufflation system was introduced, which maintains stable pneumoperitoneum while continuously evacuating smoke. Although this device has been validated in adult patients, it has not been extensively validated in the pediatric population.