Background: Patients receiving mechanical ventilation often require airway clearance and inhaled therapies. Intrapulmonary percussive ventilation (IPV) combines a high-frequency percussive ventilator with a jet nebulizer. Data on aerosol delivery efficiency of the device are scarce. We evaluated albuterol delivery efficiency while using an IPV in-line adapter under different conditions.

Methods: A pediatric lung model of invasive mechanical ventilation was used. The following independent variables were evaluated: lung condition (normal vs ARDS), ventilator mode (adaptive pressure ventilation vs pressure control), percent opening of adapter (0% vs 25% vs 50%), IPV driving pressure (25 psi vs 40 psi), IPV percussion setting (easy vs hard), and endotracheal tube (ETT) size (3.5 mm vs 5.5 mm). Albuterol delivery efficiency (mass captured in the filter expressed as percentage of loading dose) was selected as the dependent variable. Albuterol was captured on a filter at the tip of the ETT and quantified via spectrophotometry (276 nm).

Results: Albuterol delivery efficiency ranged from 0-2.89%. Median (interquartile range) and 95% CI around the median were 0.54% (0.37-1.00) and 0.50-0.63%, respectively. The coefficient of determination (R) for the model including all variables was 0.363. The 2 main contributors were percent of adapter opening (R 0.30) and IPV setting (R 0.039).

Conclusions: Albuterol delivery during invasive mechanical ventilation via in-line IPV in a pediatric lung model was inefficient. Alternative methods of delivering bronchodilators and other inhaled medications should be considered when IPV is used.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10753614PMC
http://dx.doi.org/10.4187/respcare.11311DOI Listing

Publication Analysis

Top Keywords

albuterol delivery
20
mechanical ventilation
16
delivery efficiency
16
invasive mechanical
12
pediatric lung
12
lung model
12
delivery invasive
8
ventilation in-line
8
intrapulmonary percussive
8
percussive ventilation
8

Similar Publications

Objective: To compare the efficacy of nebulized salbutamol in reducing respiratory distress in late preterm and term neonates with transient tachypnea of newborn (TTN).

Study Design: Double-blind, placebo-controlled randomized trial.

Methods: Neonates with TTN (n = 134) were allocated to nebulized salbutamol (n = 67) versus placebo (normal saline) (n = 67).

View Article and Find Full Text PDF

Comparison of Inspiration-Synchronized and Continuous Vibrating Mesh Nebulizer During Adult Invasive Mechanical Ventilation.

J Aerosol Med Pulm Drug Deliv

December 2024

Department of Cardiopulmonary Sciences, Division of Respiratory Care, Rush University, Chicago, Illinois, USA.

Aerosol delivery may be enhanced by utilizing an inspiration-synchronized nebulization mode, where nebulization occurs only during inspiration. This study aimed to compare aerosol delivery of albuterol via a prototype of an inspiration-synchronized vibrating mesh nebulizer (VMN) versus continuous VMN during invasive mechanical ventilation. A critical care ventilator equipped with a heated-wire circuit to deliver adult parameters was attached to an endotracheal tube (ETT), a collection filter, and a test lung.

View Article and Find Full Text PDF

It remains challenging to quantify lung pharmacokinetics (PK) of a drug administered and targeted to act in the lung. Exhaled breath particles (PEx), which are generated when collapsed distal airways reopen during inhalation, offer a noninvasive way to access undiluted epithelial lining fluid (ELF). Therefore, it was the aim of this study to investigate whether PK data can be derived from PEx.

View Article and Find Full Text PDF

Background: Conflicting guidance exists regarding the characterization of nebulization as an aerosol-generating procedure and subsequent risk of transmission of SARS-CoV-2 to healthcare workers.

Aim/objective: This study sought to address whether SARS-CoV-2 viral load distribution was impacted by bronchodilator delivery via nebulizer versus metered-dose inhaler (MDI).

Methods: Adults infected with COVID-19 were enrolled and received a single dose of albuterol sulfate nebulized solution (2.

View Article and Find Full Text PDF

Globally, more than 1.2 billion inhalers are purchased for asthma and chronic obstructive pulmonary disease (COPD) annually. In Australia and New Zealand, pressurized metered dose inhalers (pMDIs) are the leading delivery device prescribed and pMDI salbutamol can be purchased over the counter in Australia.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!