Technical note: 3D-printed MRI-compatible syringe pump.

Med Phys

Department of Radiology, The University of Alabama at Birmingham, Birmingham, Alabama, USA.

Published: November 2023

Background: A power injector is typically employed to deliver a contrast agent to a subject (e.g., humans, animals, phantoms) during MRI, but it is costly and cannot inject more than one subject at a time. Conventional syringe pumps housing multiple syringes are typically non-MRI compatible and, thus, should be placed outside the MRI room.

Purpose: To develop the prototype of a cost-effective, easy-to-use, reliable syringe pump that could be deployed inside a clinical MRI bore and operated on during a scan.

Methods: The MRI-compatible syringe pump (MSP) was designed using Solidworks 3D modeling software and fabricated using a Raise3D Pro2 Printer. The MSP was designed to infuse up to three syringes simultaneously. The injection speed was mainly controlled with different gear sets in the escapement unit and further adjusted by changing the effective hairspring length via a pinch pin. The MSP was evaluated with three gear sets (gear ratios: 0.20:1, 0.56:1, and 1.09:1) and 10 different effective hairspring lengths at each gear ratio. A video was recorded while operating MSP at each injection speed to calculate the volume injection rate of a 5-mL syringe (mL/s). The MSP was operated five times repeatedly at each injection speed, and the mean and standard deviation of the volume injection rate were calculated.

Results: The volume injection rates produced by three gear ratios (0.20:1, 0.56:1, and 1.09:1) were 0.209 ± 0.003 mL/s, 0.411 ± 0.002 mL/s, and 0.625 ± 0.006 mL/s, respectively, at the full hairspring length. The injection rates of gear set 1 (gear ratio: 0.20:1) decreased from 0.273 ± 0.001 mL/s to 0.245 ± 0.001 mL/s with a decrement of 0.003 mL/s for 10 different effective hairspring lengths (r = 0.997, p < 0.001). The injection rates of gear set 2 (gear ratio: 0.56:1) decreased from 0.519 ± 0.003 mL/s to 0.469 ± 0.003 mL/s with a decrement of 0.006 mL/s (r = 0.987, p < 0.001), and that of gear set 3 (gear ratio: 1.09:1) decreased from 0.779 ± 0.012 mL/s to 0.709 ± 0.005 mL/s with a decrement of 0.007 mL/s (r = 0.963, p < 0.001). The coefficient of variation in the injection rate measurement was 0.727 ± 0.346%.

Conclusions: The MSP is a portable device that can reliably deliver a liquid agent to multiple subjects inside a clinical MRI bore during a scan.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.16769DOI Listing

Publication Analysis

Top Keywords

syringe pump
12
injection speed
12
effective hairspring
12
volume injection
12
mri-compatible syringe
8
msp designed
8
gear sets
8
hairspring length
8
three gear
8
gear ratios
8

Similar Publications

Construction and experimental validation of electrochemical cells with multiple electrodes in a microfluidic channel is described. Details of the fabrication of the electrodes and polydimethylsiloxane channel using soft lithography methods are given. Calibration of the collection efficiencies and transit times between electrodes validate the use of these cells for fast electrochemical detection of soluble species.

View Article and Find Full Text PDF

Inkjet printing techniques are often used for bioprinting purposes because of their excellent printing characteristics, such as high cell viability and low apoptotic rate, contactless, commercial availability, and low cost. However, they face some disadvantages, such as the use of bioinks of low viscosity, cell damage due to shear stress caused by drop ejection and jetting velocity, as well as a narrow range of available bioinks that still challenge the inkjet printing technology. New technological solutions are required to overcome these obstacles.

View Article and Find Full Text PDF

Background: Prophylactic parenteral administration of antibiotics is strongly recommended to prevent surgical site infection (SSI). Cefoxitin is mainly administered intravenously in colorectal surgery. The current standard method for administering prophylactic antibiotics in adults is to administer a fixed dose quickly before skin incision.

View Article and Find Full Text PDF

RAFT Dispersion Polymerization of 2-Hydroxyethyl Methacrylate in Non-polar Media.

Macromolecules

December 2024

Dainton Building, Department of Chemistry, University of Sheffield, Brook Hill, Sheffield, South Yorkshire S3 7HF, U.K.

We report the reversible addition-fragmentation chain transfer (RAFT) dispersion polymerization of 2-hydroxyethyl methacrylate (HEMA) in -dodecane using a poly(lauryl methacrylate) (PLMA) precursor at 90 °C. This formulation is an example of polymerization-induced self-assembly (PISA), which leads to the formation of a colloidal dispersion of spherical PLMA-PHEMA nanoparticles at 10-20% w/w solids. PISA syntheses involving polar monomers in non-polar media have been previously reported but this particular system offers some unexpected and interesting challenges in terms of both synthesis and characterization.

View Article and Find Full Text PDF

Background: Cell concentration in body fluid is an important factor for clinical diagnosis. The traditional method involves clinicians manually counting cells under microscopes, which is labor-intensive. Automated cell concentration estimation can be achieved using flow cytometers; however, their high cost limits accessibility.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!