Background: The objective of this study was to assess the pulsatility preservation capability of the universal ventricular assist device (UVAD) when used as a biventricular assist device (BVAD). This evaluation was conducted through an in vitro experiment, utilizing a pulsatile biventricular circulatory mock loop.

Methods: Two UVAD pumps were tested in a dual setup (BVAD) in the circulatory model with the simulated conditions of left heart failure (HF), right HF, and moderate/severe biventricular HF (BHF). The total flow, aortic pulse pressure, the pulse augmentation factor (PAF), the energy-equivalent pressure (EEP), and the surplus hemodynamic energy (SHE) were observed at various pump speeds to evaluate the pulsatility.

Results: The aortic pulse pressure increased from the baseline (without pump) in all simulated hemodynamic conditions. The PAF ranged from 17%-35% in healthy, left HF, right HF, and mild BHF conditions, with the highest PAF of 90% being observed in the severe BHF condition. The EEP correlated with LVAD flow in all groups (R  = 0.87-0.97) and increased from the baseline in all cases. The SHE peaked at approximately 5-6 L/min of LVAD support and was likely to decrease at higher LVAD pump flow. The largest decrease in SHE from the baseline, 53%, was observed in the mild BHF conditions with the highest LVAD and RVAD support.

Conclusions: The UVAD successfully demonstrated the ability to preserve pulsatility in vitro, and to optimize the cardiac output, as an isolated circulatory support device option (RVAD or LVAD) and when used for BVAD support.

Download full-text PDF

Source
http://dx.doi.org/10.1111/aor.14656DOI Listing

Publication Analysis

Top Keywords

assist device
12
universal ventricular
8
ventricular assist
8
aortic pulse
8
pulse pressure
8
increased baseline
8
mild bhf
8
bhf conditions
8
conditions highest
8
lvad
5

Similar Publications

The patient's mental health from the perspective of consultation-liaison psychiatry at the stage of qualification for mechanical circulatory support implantation.

Psychiatr Pol

October 2024

Śląskie Centrum Chorób Serca w Zabrzu; Katedra i Klinika Kardiochirurgii, Transplantologii, Chirurgii Naczyniowej i Endowaskularnej, Wydział Nauk Medycznych w Zabrzu, SUM w Katowicach.

During qualification for mechanical circulatory support, the comprehensive assessment of a patient's mental state is an integral element of the overall medical evaluation. It encompasses a range of psychosocial issues, and as such provides information helpful in the selection of a suitable candidate for the required treatment, and sometimes identifies contraindications to it. It allows ensuring that the patient meets expectations regarding both mental health stability and adherence to medical recommendations.

View Article and Find Full Text PDF

We demonstrate that performing anatomical pulmonary resection by video-assisted thoracoscopic surgery without staplers or energy devices is feasible. This technique is an alternative for surgeons with limited access to expensive technologies.

View Article and Find Full Text PDF

Mechanical ventilation is the process through which breathing support is provided to patients who face inconvenience during respiration. During the pandemic, many people were suffering from lung disorders, which elevated the demand for mechanical ventilators. The handling of mechanical ventilators is to be done under the assistance of trained professionals and demands the selection of ideal parameters.

View Article and Find Full Text PDF

Ultrasound is a primary diagnostic tool commonly used to evaluate internal body structures, including organs, blood vessels, the musculoskeletal system, and fetal development. Due to challenges such as operator dependence, noise, limited field of view, difficulty in imaging through bone and air, and variability across different systems, diagnosing abnormalities in ultrasound images is particularly challenging for less experienced clinicians. The development of artificial intelligence (AI) technology could assist in the diagnosis of ultrasound images.

View Article and Find Full Text PDF

Background: Precise and accurate glenoid preparation is important for the success of shoulder arthroplasty. Despite advancements in preoperative planning software and enabling technologies, most surgeons execute the procedure manually. Patient-specific instrumentation (PSI) facilitates accurate glenoid guide pin placement for cannulated reaming; however, few commercially available systems offer depth of reaming control.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!