Aims: The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant.

Methods: This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years' follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.

Results: There were 168 patients (171 ankles) included with a mean follow-up of 2.81 years (2 to 4.6) and mean age of 63.0 years (SD 9.4). Of the ten ankles with implant failure (5.8%), six had loosening of the tibial component. In the remaining four failed implants, one was due to periprosthetic joint infection (PJI), one was due to loosening of the talar component, and two were due to loosening of both the tibial and talar components. Seven patients underwent reoperation: irrigation and debridement for superficial infection (n = 4); bone grafting for cysts (n = 2); and open reduction internal fixation (n = 1). Asymptomatic peri-implant lucency/subsidence occurred in 20.1% of ankles, with the majority involving the tibial component (n = 25). There were statistically significant improvements in PROMs in all domains.

Conclusion: Short-term results of this implant demonstrate early survival comparable to the reported survivorship of similar low-profile, non-stemmed implants. Radiological lucency occurred more commonly at the tibial component, and revisions occurred primarily due to loosening of the tibial component. Further research is needed to evaluate longer-term survivorship.

Download full-text PDF

Source
http://dx.doi.org/10.1302/0301-620X.105B10.BJJ-2023-0173.R2DOI Listing

Publication Analysis

Top Keywords

tibial component
16
total ankle
12
loosening tibial
12
patients underwent
8
vantage implant
8
implant
5
tibial
5
component
5
early outcomes
4
outcomes exactech
4

Similar Publications

Background And Purpose:  Early migration of the uncemented cruciate-sacrificing rotating platform ATTUNE and Low Contact Stress (LCS) tibial components was classified as at-risk for aseptic loosening rates exceeding 6.5% at 15 years based on recent fixation-specific migration thresholds. In this secondary report of a randomized controlled trial (RCT) we aimed to evaluate whether the 5-year migration, inducible displacement, and the clinical outcome of the ATTUNE components were comparable to those of the LCS.

View Article and Find Full Text PDF

Background: Tibial bone fractures in the malleolar regions are a major concern during the early postoperative period of total ankle replacement (TAR), affecting patient outcomes such as stability and recovery. Design, placement, and anatomic misalignment of implant components can contribute to malleolar fractures. The aim of this study is to understand the influence of implant design features, including keel, peg, stem, and bar type design, and bone-implant interfacial conditions on malleolar fracture following TAR.

View Article and Find Full Text PDF

Purpose: Double-level osteotomies (DLOs) have shown promising results for knee joint preservation, however, most ultimately progress in terms of degenerative disease resulting in conversion to total knee arthroplasty (TKA). Therefore, the purpose of this study was to examine the time to TKA conversion, long-term clinical outcomes and revision rates of patients who have undergone TKA after prior ipsilateral DLO.

Methods: Patients who underwent simultaneous or staged DLO and subsequently underwent conversion to TKA at a single academic institution from 1997 to 2022 were evaluated.

View Article and Find Full Text PDF

Background And Purpose:  This study updates 2 parallel systematic reviews and meta-analyses from 2012, which established the 1-year radiostereometric (RSA) migration thresholds for tibial components of total knee replacements (TKR) based on the risk of late revision for aseptic loosening from survival studies. The primary aim of this study was to determine the (mis)categorization rate of the 2012 thresholds using the updated review as a validation dataset. Secondary aims were evaluation of 6-month migration, mean continuous (1- to 2-year) migration, and fixation-specific thresholds for tibial component migration.

View Article and Find Full Text PDF

Intraoperatively identified patellofemoral osteoarthritis: no significant impact on short-term outcomes of medial unicompartmental knee arthroplasty.

BMC Musculoskelet Disord

January 2025

Department of Orthopaedic Surgery, The First Affiliated Hospital, Fujian Medical University, No.20, Chazhong road, Fuzhou, 350005, China.

Background And Objective: The efficacy of medial unicompartmental knee arthroplasty (UKA) in patients with intraoperatively identified patellofemoral osteoarthritis (PFOA) has been a subject of debate. This retrospective study aimed to investigate the early outcomes of UKA in patients with varying intraoperative PFOA conditions and to explore the relationship between the location of PFOA and the position of the prosthesis post-UKA. Our aim was to determine whether the presence of PFOA affects the short-term success of medial UKA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?