To address the unmet clinical need for pediatric circulatory support, we are developing an operationally versatile, hybrid, continuous-flow, total artificial heart ("Dragon Heart"). This device integrates a magnetically levitated axial and centrifugal blood pump. Here, we utilized a validated axial flow pump, and we focused on the development of the centrifugal pump. A motor was integrated to drive the centrifugal pump, achieving 50% size reduction. The motor design was simulated by finite element analysis, and pump design improvement was attained by computational fluid dynamics. A prototype centrifugal pump was constructed from biocompatible 3D printed parts for the housing and machined metal parts for the drive system. Centrifugal prototype testing was conducted using water and then bovine blood. The fully combined device ( i.e. , axial pump nested inside of the centrifugal pump) was tested to ensure proper operation. We demonstrated the hydraulic performance of the two pumps operating in tandem, and we found that the centrifugal blood pump performance was not adversely impacted by the simultaneous operation of the axial blood pump. The current iteration of this design achieved a range of operation overlapping our target range. Future design iterations will further reduce size and incorporate complete and active magnetic levitation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0000000000002043DOI Listing

Publication Analysis

Top Keywords

centrifugal pump
16
blood pump
12
pump
10
total artificial
8
artificial heart
8
centrifugal blood
8
centrifugal
7
generation development
4
development hybrid
4
hybrid continuous
4

Similar Publications

Extracorporeal Membrane Oxygenation (ECMO) is a modality of extracorporeal life support which allows temporary support in cases of cardiopulmonary failure and cardiogenic shock. This study presents a valveless pump that works by the Liebau effect as a possible pumping system in ECMO circuits, replacing the current roller and centrifugal pumps. For this purpose, a mock circulatory loop emulating the haemodynamic of the right part of the heart has been constructed.

View Article and Find Full Text PDF

In recent decades, electrokinetic handling of microparticles and biological cells found many applications ranging from biomedical diagnostics to microscale assembly. The integration of electrokinetic handling such as dielectrophoresis (DEP) greatly benefits microfluidic point-of-care systems as many modern assays require cell handling. Compared to traditional pump-driven microfluidics, typically used for DEP applications, centrifugal CD microfluidics provides the ability to consolidate various liquid handling tasks in self-contained discs under the control of a single motor.

View Article and Find Full Text PDF

Impeller radial gap is one of important parts within a blood pump, which may affect the hemodynamics and hemocompatibility. In this study, computational fluid dynamics method was performed to evaluate the impact of radial gap sizes. The volume of scalar shear stress decreased with radial gap sizes increasing.

View Article and Find Full Text PDF

Advancement of the Dragon Heart 7-Series for Pediatric Patients With Heart Failure.

Artif Organs

January 2025

BioCirc Research Laboratory, School of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, Pennsylvania, USA.

Background: Safe and effective pediatric blood pumps continue to lag far behind those developed for adults. To address this growing unmet clinical need, we are developing a hybrid, continuous-flow, magnetically levitated, pediatric total artificial heart (TAH). Our hybrid TAH design, the Dragon Heart (DH), integrates both an axial flow and centrifugal flow blood pump within a single, compact housing.

View Article and Find Full Text PDF

Progress of extracorporeal centrifugal pumps for mechanical circulatory supports.

J Artif Organs

January 2025

Department of Artificial Organs, National Cerebral and Cardiovascular Center, 6-1 Kishibe-Shinmachi, Suita, Osaka, 5648565, Japan.

This review traces the evolution of centrifugal blood pumps in mechanical circulatory support (MCS) systems. Initially met with concerns over blood damage and thrombus formation, centrifugal pumps have become crucial components in ventricular assist devices (VADs) and extracorporeal membrane oxygenation (ECMO) due to their simplified drive mechanisms and adaptability. This paper outlines three generations of centrifugal pump development: first-generation pumps with sealing components, second-generation pumps utilizing pivot bearings, and third-generation pumps employing contactless bearings.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!