[Characteristics of the Z-resolution Measurement in the Digital Breast Tomosynthesis].

Nihon Hoshasen Gijutsu Gakkai Zasshi

Department of Radiology, Division of Medical Technology, Okayama University Hospital.

Published: November 2023

Purpose: Digital breast tomosynthesis (DBT) offers a choice of image acquisition conditions. The object of observation is the entire detector in the full depth (Z-axis) direction. In this study, we investigated the effect of each acquisition condition on Z-resolution with a micro-metallic sphere (bead) as an imaged object in DBT images.

Method: DBT images were acquired in two imaging modes with different angular ranges, varying with the height of the bead, the bead placement position, and the bead diameter. The depth profile of the bead image was obtained, and the Z-resolution was defined by its full width at half maximum. The Z-resolution was compared with each acquisition condition.

Result: The Z-resolution values decreased with increasing height from the breast support and decreasing bead diameter. The Z-resolution values were larger at the edges than at the detector center.

Conclusion: This study shows that the bead diameter and the geometry of the bead and detector affect the Z-resolution measurement.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.2023-1395DOI Listing

Publication Analysis

Top Keywords

bead diameter
12
z-resolution measurement
8
digital breast
8
bead
8
z-resolution values
8
z-resolution
6
[characteristics z-resolution
4
measurement digital
4
breast tomosynthesis]
4
tomosynthesis] purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!