This article presents a prototype of a new, non-invasive, cuffless, self-calibrating blood pressure measuring device equipped with a pneumatic pressure sensor. The developed sensor has a double function: it measures the waveform of blood pressure and calibrates the device. The device was used to conduct proof-of-concept measurements on 10 volunteers. The main novelty of the device is the pneumatic pressure sensor, which works on the principle of a pneumatic nozzle flapper amplifier with negative feedback. The developed device does not require a cuff and can be used on arteries where cuff placement would be impossible (e.g., on the carotid artery). The obtained results showed that the systolic and diastolic pressure measurement errors of the proposed device did not exceed ±6.6% and ±8.1%, respectively.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10536530PMC
http://dx.doi.org/10.3390/s23187985DOI Listing

Publication Analysis

Top Keywords

blood pressure
12
pneumatic pressure
8
pressure sensor
8
device
7
pressure
6
automatic calibration
4
calibration device
4
device blood
4
pressure waveform
4
waveform measurement
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!