Development of an ID-LC-MS/MS method using targeted proteomics for quantifying cardiac troponin I in human serum.

Clin Proteomics

Department of Molecular Biology and Genetics, Faculty of Science and Letters, Istanbul Technical University, Istanbul, Turkey.

Published: September 2023

AI Article Synopsis

Article Abstract

Background: Cardiac troponin is a complex protein consisting of the three subunits I, T and C located in heart muscle cells. When the heart muscle is damaged, it is released into the blood and can be detected. Cardiac troponin I (cTnI) is considered the most reliable and widely accepted test for detecting and confirming acute myocardial infarction. However, there is no current standardization between the commercial assays for cTnI quantification. Our work aims to create a measurement procedure that is traceable to the International System of Units for accurately measuring cardiac cTnI levels in serum samples from patients.

Methods: The workflow begins with immobilizing anti-cTnI antibodies onto magnetic nanoparticles to form complexes. These complexes are used to isolate cTnI from serum. Next, trypsin is used to enzymatically digest the isolated cTnI. Finally, the measurement of multiple cTnI peptides is done simultaneously using isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS).

Results: The maximum antibody immobilization was achieved by combining 1 mg of nanoparticles with 100 μg of antibody, resulting in an average of 59.2 ± 5.7 μg/mg of immobilized antibody. Subsequently, the anti-cTnI-magnetic nanoparticle complex was utilized to develop and validate a method for quantifying cTnI in human serum using ID-LC-MS/MS and a protein calibration approach. The analytical method was assessed regarding linearity and recovery. The developed method enables the quantification of cTnI from 0.7 to 24 μg/L (R > 0.996). The limit of quantification was 1.8 μg/L and the limit of detection was 0.6 μg/L. Intermediate precision was ≤ 9.6% and repeatability was 2.0-8.7% for all quality control materials. The accuracy of the analyzed quality control materials was between 90 and 110%. Total measurement uncertainties for target value assignment (n = 6) were found to be ≤ 12.5% for all levels.

Conclusions: The analytical method demonstrated high analytical performance in accurately quantifying cardiac troponin I levels in human serum. The proposed analytical method has the potential to facilitate the harmonization of cTnI results between clinical laboratories, assign target values to secondary certified reference materials and support reliable measurement of cTnI.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10536812PMC
http://dx.doi.org/10.1186/s12014-023-09430-zDOI Listing

Publication Analysis

Top Keywords

cardiac troponin
16
human serum
12
analytical method
12
ctni
10
quantifying cardiac
8
heart muscle
8
quality control
8
control materials
8
method
6
cardiac
5

Similar Publications

Intra-Mesopore Immunoassay Based on Core-Shell Structured Magnetic Hierarchically Porous ZIFs.

ACS Sens

January 2025

Key Laboratory for Ultrafine Materials of Ministry of Education, School of Materials Science and Engineering, East China University of Science and Technology, Shanghai 200237, China.

It is crucial yet challenging to sensitively quantify low-abundance biomarkers in blood for early screening and diagnosis of various diseases. Herein, an analytical model of intra-mesopore immunoassay (IMIA) was proposed, which was competent to examine various biomarkers at the femtomolar level. The success is rooted in the design of an innovative superparamagnetic core-shell structure with FeO nanoparticles (NPs) at the core and hierarchically porous zeolitic imidazolate frameworks as a shell (FeO@HPZIF-8), achieved through a soft-template directed self-assembly coupled with confinement growth mechanism.

View Article and Find Full Text PDF

Background: Glucagon-like peptide-1 (GLP-1) agonists are an existing treatment option for patients with insulin-resistant states, which elicit further pleiotropic effects related to immune cell recruitment and vascular inflammation. GLP-1 agonists downregulate the cluster of differentiation 147 (CD147) receptor, one of several receptors for the SARS-CoV-2 spike protein that mediate viral infection of host cells.

Methods: We conducted an open-label prospective safety and tolerability study including biomarker responses of the GLP-1 agonist Liraglutide, administered for 5 days as an add-on therapy to the standard of care within 48 h of presentation in a cohort of 13 patients hospitalized with COVID-19 pneumonia.

View Article and Find Full Text PDF

Background: Cardiac myosin binding protein C (cMyC) is an emerging new biomarker of myocardial injury rising earlier and cleared faster than cardiac troponins. It has discriminatory power similar to high-sensitive troponins in diagnosing myocardial infarction in patients presenting with chest pain. It is also associated with outcome in patients with acute heart failure.

View Article and Find Full Text PDF

Background: Myocardial injury is common after brain injury; however, few studies have reported serial cardiac troponin (cTn) measurements to distinguish whether the myocardial injury is acute or chronic. The fourth Universal Definition of Myocardial Infarction introduced for the first time the criteria for acute myocardial injury (AMI). We aimed to investigate the prevalence and prognostic implications of AMI in primary intracerebral hemorrhage.

View Article and Find Full Text PDF

Aims: A simplified version of the history, electrocardiogram, age, risk factors, troponin (HEART) score, excluding troponin, has been proposed to rule-out major adverse cardiac events (MACEs). Computerized history taking (CHT) provides a systematic and automated method to obtain information necessary to calculate the HEAR score. We aimed to evaluate the efficacy and diagnostic accuracy of CHT in calculating the HEAR score for predicting MACE.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!