Background: Left atrial appendage (LAA) closure (LAAC) procedures are conventionally performed using empirical fluoroscopic viewing angles. However, because the LAA is a highly variable anatomical structure, these angles cannot depict the LAA in the optimal position. The present study aimed to assess the efficiency of using a novel optimal fluoroscopic projection angle (OPA) for LAAC and to validate its feasibility.

Methods: The OPAs of the derivation cohort were acquired using cardiac computed tomography angiography (CCTA) to assess its superiority for depicting LAA depth versus traditional working angles (TAs) of RAO 30°, CAU 20°. The practicability of OPA-guided LAAC was demonstrated by comparison between clinical data from the validation cohort and those from a propensity-score matched (PSM) control group, as well as randomized controlled studies investigating LAAC.

Results: Of 705 patients in the derivation cohort, the median OPA was RAO 46°, CAU 31°. Compared with TA, the OPA depicted a longer mean (±SD) LAA depth (5.1 ± 4.4) mm and a larger orifice diameter (1.1 ± 1.1 mm), (P < 0.0001 for both). All 38 OPA-guided LAACs were successful, with a shorter mean procedure duration (42.9 ± 12.3 min versus [vs.] 107.2 ± 41.5 min; P < 0.0001) and reduced device consumption (1.08 vs. 1.5 per case), compared with the PSM control group. At the 3-month follow-up, the incidence of peri-device leak was 52.6% (20/38) detected by CCTA, with a mean leakage of 1.6 ± 0.8 mm.

Conclusion: By unfolding the LAA depth and orifice diameter for a better view, OPA demonstrated the potential to optimize LAAC procedural efficiency, although further larger-scale studies are required to confirm this.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hjc.2023.09.009DOI Listing

Publication Analysis

Top Keywords

optimal fluoroscopic
8
fluoroscopic projection
8
projection angle
8
computed tomography
8
tomography angiography
8
left atrial
8
atrial appendage
8
derivation cohort
8
laa depth
8
laa
5

Similar Publications

Introduction: Intertrochanteric fractures are common in older adults and pose significant challenges in terms of morbidity and mortality. Accurate reduction and optimal implant positioning during operative stabilisation of these fractures reduce the rates of complications and reoperations while improving functional outcomes in this population. This study aimed to assess the effects of a structured educational intervention on the radiographic outcomes, reduction quality, and revision rates of intertrochanteric fractures.

View Article and Find Full Text PDF

Objective: There are many advantages to stabilize the posterior pelvic ring injuries with a transiliac-transsacral (TITS) screw percutaneously. To identify the correct entry point and insert a guidewire accurately for a TITS screw, we propose a method of specifying the optimal entry point, and introduce a technique of enabling freehand placement of a guidewire with fluoroscopic guidance.

Methods: In this retrospective study, 116 patients who underwent pelvic CT scans and pelvic lateral radiographs at our institution from January 2020 to April 2022 were enrolled.

View Article and Find Full Text PDF

All Enabling Technology Is Not Created Equal: Comparing Outcomes of Computer-Assisted Fluoroscopic Navigation Versus Robotic-Assisted Total Hip Arthroplasty.

J Am Acad Orthop Surg Glob Res Rev

December 2024

From the Connecticut Orthopaedics, Fairfield, CT (Dr. Bernstein); the Epidemiology and Real-World Data Sciences, Johnson & Johnson MedTech, New Brunswick, NJ (Dr. Gupta and Ruppenkamp); the Global Health Economics and Market Access, Johnson & Johnson MedTech, Raynham, MA (Dr. Kabiri and Goldstein); and the Medical Affairs, Johnson & Johnson MedTech, Palm Beach Gardens, FL (Dr. Diaz).

Background: Computer-assisted fluoroscopic navigation and robotic technologies aim to optimize implant placement and alignment in primary total hip arthroplasty (THA) to improve patient outcomes. This study uses a retrospective hospital billing database covering 1,300 hospitals to compare the clinical and economic effect of these technologies.

Methods: The study compared patients undergoing THA with robotic versus computer-assisted fluoroscopic navigation technologies between January 1, 2016, and September 30, 2021, using the Premier Healthcare Database.

View Article and Find Full Text PDF

Background: The standard of care for placement of a central venous catheter (CVC) includes a real-time ultrasound (US)-guided technique. We describe a rare case in which the guidewire penetrated the posterior wall of the vessel, forming a knot, which precluded simple removal. This occurred despite the procedure being performed under real-time US guidance.

View Article and Find Full Text PDF

A 78-year-old female with a remote history of L3-4 decompression and fusion presented with several months of low back and radicular leg pain. MRI revealed moderate L2-L3 spinal canal stenosis, ligamentum flavum infolding, moderate bilateral foraminal stenosis, and a grade I retrolisthesis. A right sided L2-L3 TFESI was performed using multiplanar fluoroscopic imaging with a subpedicular supraneural approach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!