Objective: Concerns regarding patient safety and image quality have made the use of knee-spanning external fixators in MRI a challenging clinical scenario. The purpose of our study was to poll practicing musculoskeletal radiologists on their personal experiences regarding the use of knee-spanning external fixators in MRI in an effort to consolidate practice trends for the radiologists' benefit.

Methods: A 27-item survey was created to address the institutional use, safety, adverse events, quality, and perspectives of the radiologist related to MRI of an externally fixated knee. The survey was distributed to 1739 members of the Society of Skeletal Radiology.

Results: A total of 72 members of the Society of Skeletal Radiology completed the survey. Most notably, 40 of 72 (55.56%) respondents are permitted to place a knee-spanning external fixator inside the MR bore at their institution, while19 of 72 (26.39%) respondents are not permitted to do so. Fourteen of 32 (43.75%) respondents have institutional guidelines for safely performing an MRI of an externally fixated knee. Twenty-five of 32 (78.13%) respondents are comfortable permitting an MRI of an externally fixated knee.

Conclusion: We found a general lack of consensus regarding the decision to scan a patient with a knee-spanning external fixator in MRI. Many institutions lack safety guidelines, and providers rely upon a heterogeneous breadth of resources for safety information. A re-examination of the FDA device labeling nomenclature and expectations of the individual manufacturers may be needed to bridge this gap and help direct management decisions placed upon the provider.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00256-023-04445-xDOI Listing

Publication Analysis

Top Keywords

knee-spanning external
20
external fixators
12
fixators mri
12
society skeletal
12
mri externally
12
externally fixated
12
patient safety
8
safety image
8
image quality
8
quality knee-spanning
8

Similar Publications

Background: Distal femur infected non-union management is very challenging especially when there is late presentation and after multiple intervention. Various methods have been described and studied in the literature for infected non-union of femur which includes bone transport by fixators (ring fixators/monolateral frames) and by Masquelet method.

Objectives: To study the clinical and radiological outcomes and study the effect on quality of life by patient-reported outcomes (EQ5D5L) and VAS (Visual Analog Scale) score in patients of infected non-union of distal femur managed with combination of Masquelet method and Ilizarov ring fixator.

View Article and Find Full Text PDF

Background: The evolution of warfare has resulted in a surge of high-energy blast injuries predominantly involving the lower extremities. Once thought to impact only forward-deployed military, such mechanisms of injury are becoming a harsh reality even in civilian territory. Proficiency in surgical techniques for extremity damage control is vital for surgeons, regardless of specialty.

View Article and Find Full Text PDF
Article Synopsis
  • Floating hip injuries, characterized by a combination of hip dislocation and fractures, are rare and challenging to treat, often resulting in poor outcomes.
  • A case study of a 35-year-old man showed successful management of his hip and femur fractures using a combination of external fixation and surgical techniques, leading to gradual recovery.
  • The conclusion emphasizes that careful staged management can improve outcomes and minimize complications in patients with complex floating hip injuries.
View Article and Find Full Text PDF

After emergent assessment of potentially limb-threatening injuries in knee dislocation or multi-ligament knee injury patients, magnetic resonance imaging is necessary to visualize ligamentous structures and plan for soft tissue repair. However, the application of a knee-spanning external fixator may introduce artifact and reduce overall image quality, which can limit the evaluation of soft tissue injury. As a result, the utility of MRI in the context of a knee-spanning external fixator has been called into question.

View Article and Find Full Text PDF

Objective: Concerns regarding patient safety and image quality have made the use of knee-spanning external fixators in MRI a challenging clinical scenario. The purpose of our study was to poll practicing musculoskeletal radiologists on their personal experiences regarding the use of knee-spanning external fixators in MRI in an effort to consolidate practice trends for the radiologists' benefit.

Methods: A 27-item survey was created to address the institutional use, safety, adverse events, quality, and perspectives of the radiologist related to MRI of an externally fixated knee.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!