We set out to determine the performance of the Testi Technologies enzymatic assay saliva ethanol test strips of three different detection levels: 0 g/L, 0.2 g/L and 0.5 g/L, using as the reference method a gas chromatography analyser (GC). Alcohol levels were measured in 104 volunteers at up to three points in time, using up to three test strips per measurement, while gathering blood samples and breathalyser readings in parallel. The plasma alcohol concentrations (PAC) were determined from the plasma samples using GC. The qualitative results of the test strips were compared to the quantitative results from the reference method, as well as the breathalyser readings, and the amount of true and false positive and true and false negative results were classified using predetermined cut-off levels. The best performing test strips were the 0 g/L and the 0.2 g/L strips. The 0 g/L strips had a sensitivity and specificity of 1.00, as false negatives and false positives were not detected. The 0.2 g/L strips had a sensitivity and specificity [95% confidence interval (CI)] of 0.98 (0.96 - 1.00) and 0.83 (0.62 - 1.00) respectively, an accuracy of 0.97 (0.95 - 0.99), and a diagnostic odds ratio of 205.00 (35.33 - 1189.66). The test strips perform their intended purpose of screening for alcohol consumption well, with their great sensitivity as a defining property compared to other testing methods. For them to be able to be implemented in a clinical setting however, further refinement of the tests' characteristics would be required.

Download full-text PDF

Source
http://dx.doi.org/10.1080/00365513.2023.2255970DOI Listing

Publication Analysis

Top Keywords

test strips
24
strips
9
gas chromatography
8
0 g/l 02 g/l
8
reference method
8
breathalyser readings
8
true false
8
strips 0 g/l
8
02 g/l strips
8
strips sensitivity
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!