Background: Failure to achieve fixation of the glenoid baseplate will lead to clinical failure. The fixation of the baseplate to the scapula must be able to withstand sufficient shear forces to allow bony ingrowth. The importance of compression to neutralize the forces at the baseplate-bone interface has been assumed to be critical in limiting excessive micromotion. The purpose of this study is to determine the effect of compression on implant stability with different baseplate designs.
Methods: Various baseplate designs (1-piece monolithic central screw [1P], 2-piece locking central screw [2PL], and 2-piece nonlocking center screw [2PNL]) were investigated at 3 different compressive forces (high [810 N], medium [640 N], and low [530 N]). Synthetic bone cylinders were instrumented, and peripheral screws were used in all models. The combination of 1 locking and 3 nonlocking peripheral screw fixation was selected as worst-case scenario. Dynamic testing protocol followed the ASTM F2028-17 standard. The baseplate micromotion at high compression was compared to low compression. Additionally, the baseplate micromotion for each design was compared at baseline (first 50 cycles) and at 10,000 cycles for the 3 different compressive forces where motion above 150 μm was defined as failure.
Results: Baseplate micromotion was found to negatively correlate with compression (r = -0.83, P < .0001). At baseline, all baseplate designs were considered stable, regardless of compression. With high compression, average micromotion at the glenoid baseplate-bone interface remained below the 150-μm threshold for all baseplate designs at 10,000 cycles (1P: 50 ± 10 μm; 2PL: 78 ± 32 μm; 2PNL: 79 ± 8 μm; P = .060). With medium compression, average micromotion at 10,000 cycles for all 3 designs remained below the 150-μm threshold (1P: 88 ± 22 μm; 2PL: 132 ± 26 μm; 2PNL: 107 ± 39 μm). The 2PL design had the highest amount of micromotion (P = .013). With low compression, both 2-piece designs had an average micromotion above the 150-μm threshold whereas the 1-piece design did not (1P: 133 ± 35 μm; 2PL: 183 ± 21 μm; 2PNL: 166 ± 39 μm). The 2PL design had significantly higher micromotion when compared to 1P design (P = .041).
Discussion: The stability of a central screw baseplate correlates with the amount of compression obtained and is affected by implant design. For the same amount of compression, more micromotion is observed in a 2-piece design than a 1-piece design.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jse.2023.07.043 | DOI Listing |
Bioengineering (Basel)
January 2025
Department of Orthopedic Research, Arthrex, 81249 Munich, Germany.
Objective: This study evaluated the effects of bony increased offset (BIO) and metallic augments (MAs) on primary reverse shoulder arthroplasty (RSA) baseplate stability in cadaveric specimens with variable bone densities.
Methods: Thirty cadaveric specimens were analyzed in an imaging and biomechanical investigation. Computed tomography (CT) scans allowed for preoperative RSA planning and bone density analysis.
J Biomech Eng
January 2025
Center for Orthopaedic Biomechanics, University of Denver, Room 427, 2155 E. Wesley Avenue, Denver, CO 80208.
Micromotion exceeding 150 μm at the implant-bone interface may prevent bone formation and limit fixation after cementless knee arthroplasty. Understanding the critical parameters impacting micromotion is required for optimal implant design and clinical performance. However, few studies have focused on unicompartmental knee arthroplasty (UKA).
View Article and Find Full Text PDFJ Shoulder Elbow Surg
February 2025
Department of Orthopaedic Surgery, New England Baptist Hospital, Boston, MA, USA; Boston Sports and Shoulder Center Research Foundation, Waltham, MA, USA. Electronic address:
Background: Augmented baseplates can be effective at addressing eccentric glenoid wear in reverse total shoulder arthroplasty. However, these implants often come in a limited number of predetermined shapes that require additional reaming to ensure adequate glenoid seating. This typically involves complex instrumentation and can have a negative impact on implant stability.
View Article and Find Full Text PDFKnee
August 2024
Biomatlab, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway; Department of Clinical Medicine, University of Bergen, Bergen, Norway.
Background: Approximately 5% of primary total knee arthroplasty patients require revision within 10 years, often due to distal component loosening. Application of a thin layer of PMMA cement as precoating on the tibial component aims to prevent aseptic loosening. This study investigates the impact of precoating and fat contamination on tibial baseplate stability.
View Article and Find Full Text PDFSemin Arthroplasty
December 2023
Department of Bioengineering, Clemson University, Clemson, SC, USA.
Background: Posterior glenoid bone loss is frequently observed in patients with osteoarthritis undergoing reverse total shoulder arthroplasty. Glenoid bone loss can reduce the baseplate back support area and the number of screws for fixation. The purpose of this study is to determine how initial baseplate fixation is affected by biomechanical factors introduced by glenoid bone loss such as reduced baseplate back support area and reduced screw number using three-dimensional finite element analysis.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!