Purpose: To evaluate coronary artery integrity after very high radiation doses from intravascular brachytherapy (IVBT) in the setting of source asymmetry.

Methods: Ten patients treated for right coronary artery (RCA) in-stent restenosis (ISR) between 2017 and 2021 and for whom follow-up angiograms were available were identified from departmental records. Procedural angiograms, taken to document source position, were used to estimate vascular wall doses. The 2.5 mm proximal source marker was used to estimate the distance from source center to the media and adventitia. Distances were converted to dose (Gy) using the manufacturers' dose fall-off table, measured in water. Follow-up films were scrutinized for any sign of late vascular damage.

Results: The average minimal distance from catheter center to the adjacent media and the adventitia was 0.9 mm (±0.2) mm and 1.4 mm (±0.2), respectively. The average maximum media and adventitial doses adjacent to the source were 75 Gy (±26) and 39 Gy (±14), respectively. Follow-up angiograms were available from 0.6 years to 3.9 years following IVBT (median: 1.6 years). No IVBT-treated vascular segment showed signs of degeneration, dissection or aneurysm.

Conclusion: IVBT vascular wall doses are frequently far higher than prescribed. The lack of complications in this unselected group of patients gives a modicum of reassurance that raising the prescription dose is unlikely to lead to a sudden appearance of complications.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.carrev.2023.08.014DOI Listing

Publication Analysis

Top Keywords

radiation doses
8
doses intravascular
8
intravascular brachytherapy
8
coronary artery
8
follow-up angiograms
8
vascular wall
8
wall doses
8
media adventitia
8
doses
5
source
5

Similar Publications

We report on a procedure for extracting the SPICE model parameters of a RADFET sensor with a dielectric HfO/SiO double-layer. RADFETs, traditionally fabricated as PMOS transistors with SiO, are enhanced by incorporating high-k dielectric materials such as HfO to reduce oxide thickness in modern radiation sensors. The fabrication steps of the sensor are outlined, and model parameters, including the threshold voltage and transconductance, are extracted based on experimental data.

View Article and Find Full Text PDF

Fiber Bragg gratings (FBGs) inscribed by UV light and different femtosecond laser techniques (phase mask, point-by-point, and plane-by-plane) were exposed-in several irradiation cycles-to accumulated high doses of gamma rays (up to 124 MGy) and neutron fluence (8.7 × 10/cm) in a research-grade nuclear reactor. The FBG peak wavelengths were measured continuously in order to monitor radiation-induced shifts.

View Article and Find Full Text PDF

With technological development, ionizing radiation has found applications in numerous occupations. However, the determination and quantification of the damage resulting from exposure to it remains rather unclear, along with the damage to particular organs. The aim of this systematic review was to investigate the relationship between low-dose ionizing radiation (LDIR) in exposed workers and possible functional changes and cancer development in the thyroid gland.

View Article and Find Full Text PDF

The melon fly, , poses a severe threat to the country's agricultural productivity, particularly in the cultivation of cucurbitaceous crops. This study was conducted to determine the ideal irradiation dose to be used to set up a Sterile Insect Technique (SIT)-based strategy to control outbreaks in Sri Lanka. A colony was established and maintained under standard laboratory conditions.

View Article and Find Full Text PDF

Radiation therapy (RT) is the cornerstone treatment for prostate cancer; however, it frequently induces gastrointestinal and genitourinary toxicities that substantially diminish the patients' quality of life. While many individuals experience transient side effects, a subset endures persistent, long-term complications. A promising strategy to mitigate these toxicities involves enhancing tumor radiosensitivity, potentially allowing for lower radiation doses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!