Background: To evaluate the possible advantages of a simple spinal cord (SC) dose-limiting three-dimensional conformal radiotherapy (3D-CRT) technique in comparison to conventional two-dimensional (2D) techniques and other 3D-CRT techniques for spinal bone irradiation.

Methods: For 41 spinal target volumes, seven different techniques were evaluated, using a standard schedule of 30 Gy in 10 fractions. The SC dose-limiting 3D-CRT technique 1F2S-18MV using a single posterior field (F) supplemented by two anterior segment fields (S) and 18-MV photon beams was compared to two conventional 2D techniques (a single posterior field, PA, and two opposed anterior-posterior fields, APPA), three other 3D-CRT techniques (a single posterior field supplemented by four segment fields, 1F4S; two wedged fields, WD, and the SC dose-limiting variant using 6 MV, 1F2S-6MV) along with the original clinically applied plans.

Results: 1F2S-18MV demonstrated notably better results for all target volume parameters compared to the conventional 2D techniques (p < 0.001). Limitation of the SC dose was significantly superior with 1F2S-18MV in comparison to PA and APPA (SC Dmean: 28.9 ± 0.4  vs. 30.1 ± 0.6 Gy and 30.1 ± 0.4 Gy; SC Dmax: 30.9 ± 0.7  vs. 32.5 ± 1.0 Gy and 31.8 ± 0.7 Gy; SC D1cm : 30.1 ± 0.6  vs. 31.7 ± 0.9 Gy and 31.1 ± 0.6 Gy; p < 0.001). Likewise, lower mean SC doses with 1F2S-18MV were observed in comparison to the more treatment time-consuming 3D-CRT techniques (1F4S, WD) and the original plans without relevant compromises on the dose homogeneity in the target volume and the dose exposure to the other OARs.

Conclusion: In treatment planning of spinal metastases, simple variants of 3D-CRT-techniques like 1F2S-18MV can offer a significant dose limitation to the SC while providing a sufficient dose coverage of the target volume. Especially in patients with favorable life expectancy and potential need for re-irradiation, such SC dose-limiting 3D-CRT techniques may be a reasonable approach.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10562034PMC
http://dx.doi.org/10.1002/acm2.14042DOI Listing

Publication Analysis

Top Keywords

3d-crt technique
12
single posterior
12
posterior field
12
spinal cord
8
cord dose-limiting
8
dose-limiting 3d-crt
8
3d-crt techniques
8
field supplemented
8
segment fields
8
compared conventional
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!