Effect of vertical pump position on start-up fluid delivery of syringe pumps used for microinfusion.

Paediatr Anaesth

Deutsches Herzzentrum der Charité (DHZC) Medicine, Department of Cardiac Anesthesiology and Intensive Care, Berlin, Germany.

Published: December 2023

Background: Connection and opening a syringe infusion pump to a central venous line can lead to acute anterograde or retrograde fluid shifts depending on the level of central venous pressure. This may lead to bolus events or to prolonged lag times of intravenous drug delivery, being particularly relevant when administering vasoactive or inotropic drugs in critically ill patients using microinfusion. The aim of this study was to assess the effect of syringe pump positioning at different vertical heights on start-up fluid delivery before versus after purging and connection the pump to the central venous catheter.

Methods: This in vitro study measured ante- and retrograde infusion volumes delivered to the central venous line after starting the syringe pump at a set infusion rate of 1 mL/h. In setup one, the pump was first positioned to vertical levels of +43 cm or -43 cm and then purged and connected to a central venous catheter. In setup two, the pump was first purged and connected at zero level and secondarily positioned to a vertical level of +43 cm or -43 cm. Central venous pressure was adjusted to 10 mmHg in both setups.

Results: Positioning of the pump prior to purging and connection to the central venous catheter resulted in a better start-up performance with delivered fluid closer to programmed and expected infusion volumes when compared to the pump first purged, connected, and then positioned. Significant backflow volumes were observed with the pump purged and connected first and then positioned below zero level. No backflow was measured with the pump positioned first below zero level and then purged and connected.

Conclusions: Syringe infusion pump assemblies should be positioned prior to purging and connection to a central venous catheter line when starting a new drug, particularly when administering highly concentrated vasoactive or inotropic drugs delivered at low flow rates.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pan.14750DOI Listing

Publication Analysis

Top Keywords

central venous
32
purged connected
16
purging connection
12
venous catheter
12
pump purged
12
pump
11
start-up fluid
8
fluid delivery
8
syringe infusion
8
infusion pump
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!