Objectives: Three-dimensional cone beam computed tomography (CBCT) imaging can be considered, especially in patients with complicated peri-implantitis (PI). Artifacts induced by dense materials are the drawback of CBCT imaging and the peri-implant bone condition may not be assessed reliably because the artifacts are present in the same area. This pilot study investigates the performance of the artifact reduction algorithm (ARA) of the Planmeca Viso G7 CBCT device (Planmeca, Helsinki, Finland) with three different implant materials and imaging parameters.

Methods: Three pairs of dental implants consisting of titanium, zirconia, and fiber reinforced composite (FRC) were set into a pig mandible. A vertical defect simulating peri-implantitis bone loss was made on the buccal side of one of each implant. The defect was identified and measured by two observers and compared to the actual dimensions. In addition, the bone structure and the marginal cortex visibility between the implants were estimated visually.

Results: The bone defect and its dimensions with the zirconia implant could not be identified in any image with or without the metal artifact reduction algorithm. The bone defect of titanium and FRC implants were identified with all three imaging parameters or even without ARA. The interobserver agreement between the two observers was almost perfect for all categories analyzed.

Conclusion: Peri-implantitis defect of the zirconia implant and the peri-implant bone structure of the zirconia implants cannot be recognized reliably with any ARA levels, or any imaging parameters used with the Planmeca Viso G7. The need for ARA when imaging the peri-implant bone condition of the titanium and FRC implants may be unnecessary.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10968758PMC
http://dx.doi.org/10.1259/dmfr.20230252DOI Listing

Publication Analysis

Top Keywords

peri-implant bone
16
artifact reduction
12
zirconia implants
8
bone
8
cone beam
8
beam computed
8
computed tomography
8
tomography cbct
8
pilot study
8
cbct imaging
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!