Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: In order to prevent magnetic materials from being brought into the magnetic resonance imaging (MRI) examination room, many facilities have metal detectors, etc., but there are various types of equipment with different performance and characteristics. The purpose of this study was to evaluate each detector in actual clinical practice.
Methods: At multiple facilities, gate-type magnetic detectors, pole-type magnetic detectors, handy-type magnetic detectors, and handy-type metal detectors were used to identify 9 types of objects that may be brought into the MRI examination room. We performed evaluation of detection distance measurement assuming actual operation.
Results: The gate type was only able to detect objects with strong magnetism. With the pole type, the closer the measurement distance was to the pole, the more objects could be detected, and the lower the pole, the shorter the detection distance. With the handy type, there were many objects that could be detected when the device and the object were brought into close contact.
Conclusion: The detectability of the instruments varied depending on the size and type of the object. It is important to understand the characteristics of each device and use it according to the purpose in carrying-in confirmation before the examination.