Background: Radiochromic material such as lithium pentacosa-10,12-diynoate (LiPCDA) has been suggested as the radiation-sensitive material for real-time in vivo fiber-optic dosimetry. In this configuration, micron-thick radiochromic coating would measure the absorbed dose, where a major challenge is the uncertainty in the active material thickness, necessitating calibration. A homogeneously incorporated inert infrared (IR) dye, which must also be stable in ambient conditions and against radiolysis, can be added to the radiochromic film to enable optical calibration.

Purpose: This study investigates four commercial cyanine-based dyes (IR-783, IR-806, IR-868, and IR-880) for use as an optical calibrant in fiber-optic radiochromic dosimeters.

Methods: All dyes were dissolved in water to confirm solubility. IR-783 and IR-806 were dissolved in 10% w/w gelatin solution and coated onto a polyester substrate, which were then sandwiched between two layers of adhesives forming IR-783 and IR-806 films. A second batch of IR dyes in gelatin incorporated the LiPCDA, and was coated onto substrate and sandwiched between adhesive to form IR dye + LiPCDA films. The absorbance spectra of the films were measured periodically (176 and 102 days for IR-dye films, and IR dye + LiPCDA, respectively). The average percentage absorbance, normalized to day 1, was fit to either a single or a double exponential decay model to calculate the spectral stability lifetime (τ , τ ). Films were irradiated using a 6 MV LINAC beam with a standard setup of 100 source to axis distance (SAD), 10 cm × 10 cm field size and 1.5 cm depth. The change in absorbance of the IR-dye + LiPCDA films were measured after they were irradiated to 1, 2, 5, 10, and 20 Gy at 3 Gy/min.

Results: Only IR-783 and IR-806 were sufficiently water soluble. In gelatin matrix, these dyes demonstrated a decrease in absorbance with time for IR-783 and IR-806 dyes, with IR-783 films having an average τ  = 73 ± 7 days and IR-806 films τ  = 7 ± 3 days. When combined with LiPCDA, IR-806 degraded, losing its original peak at ∼820 nm. Similarly, IR-783, combined with LiPCDA, showed signs of degradation; however, its original absorbance peak was still observed at ∼800 nm. In the IR-783 + LiPCDA films, the IR-783 dye had a τ = 4 ± 1 days, an order of magnitude faster than the IR-783 with no LiPCDA films. When exposed to x-ray irradiation, the IR-783 dye in the IR-783 + LiPCDA films showed no change in absorbance with increasing absorbed dose. In contrast, the LiPCDA in the films responded as expected, increasing in optical density with increased absorbed dose.

Conclusions: IR-783 and IR-806 dyes were observed to degrade over time following exponential decay curves. IR-806 could not be combined with the LiPCDA without degrading. The combination of IR-783 with LiPCDA demonstrated single exponential decay behavior at a comparatively faster rate than films that did not have LiPCDA. IR-783 was insensitive to ionizing radiation and thus may be suitable for thickness correction, but an alternative manufacturing procedure may need to be developed.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.16674DOI Listing

Publication Analysis

Top Keywords

ir-783 ir-806
24
films
14
ir-783
13
exponential decay
12
combined lipcda
12
lipcda
9
ir-806
9
absorbed dose
8
dyes ir-783
8
substrate sandwiched
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!