Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: Urine output (UO) is an important intraoperative parameter that is not yet electronically monitored. We compared an automatic urinometer (AU) based on a smart scale with a manual urinometer (MU).
Patients And Methods: This prospective study investigated the hourly UO of 35 preoperative patients with an indwelling urinary catheter using AU, MU, and cylinder measurements. Data were analyzed using the Bland-Altman method. A questionnaire related to the use of the AU was completed by medical staff (n=25).
Results: Compared to the cylinder measurements, the differences in measurements by the AU and the MU were -6.31 ± 15.03 mL/h (p=0.018) and 20.26 ± 26.81 mL/h (p=0.001), respectively. The r values for the comparison of cylinder measurements with AU and MU values were 0.985 (p<0.001) and 0.968 (p<0.001), respectively. Bland-Altman analyses showed that cylinder measurements had better agreement with the AU measurements than with the MU measurements. Also, the medical staff reported that the use of the AU was easier to learn than the use of the MU (p<0.001).
Conclusions: Compared to the MU values, AU values were noninferior; they had significantly less bias and temporal deviation. Additionally, the medical staff reported that the use of the AU was easier to learn than the use of the MU.