Background/purpose: To assess usefulness of phacofragmentation probe in the removal of silicone oil (SO) adherent to intraocular lenses (IOLs) during a SO removal procedure.
Methods: A 20-gauge (G) phacofragmentation probe was used to emulsify and remove SO adherent to IOLs during 23-G vitrectomy for SO removal. The superotemporal 23-G cannula was removed, and 20-G sclerotomy was made by using a 20-G MVR blade.
Results: In all cases used, SO adherent to the posterior surface of posterior capsule IOLs and anterior chamber IOLs was easily removed using a phacofragmentation probe. No intraoperative complications or IOL damage were noted.
Conclusion: We describe a simple, quick and reliable method of SO removal from all types of IOLs by using a phacofragmentation probe during SO removal surgery. This simple method can prevent IOL removal and replacement due to adherent SO bubbles in pseudophakic eyes that were previously filled with SO.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/IAE.0000000000003199 | DOI Listing |
Retina
September 2023
Retina and Uveitis Service, Havener Eye Institute, the Ohio State University Wexner Medical Center, Columbus, Ohio.
Background/purpose: To assess usefulness of phacofragmentation probe in the removal of silicone oil (SO) adherent to intraocular lenses (IOLs) during a SO removal procedure.
Methods: A 20-gauge (G) phacofragmentation probe was used to emulsify and remove SO adherent to IOLs during 23-G vitrectomy for SO removal. The superotemporal 23-G cannula was removed, and 20-G sclerotomy was made by using a 20-G MVR blade.
Ophthalmic Surg Lasers Imaging
February 2011
Asociación Para Evitar la Ceguera en México, Hospital Dr. Luis Sánchez Bulnes, Coyoacán, México.
The authors describe a modified technique for extraction of a dislocated crystalline lens or large and hard lens fragments displaced into the vitreous cavity. An ophthalmic viscosurgical device was used around and on top of the heavy liquid bubble to keep the lens centered and less mobile for easier phacofragmentation or removal by the cutting-suction probe. This technique was used in 10 eyes, including cases of traumatic dislocation and Marfan syndrome, with excellent anatomical results and no complications that could be attributed to the surgical technique.
View Article and Find Full Text PDFA new intraocular 20-gauge (0.89-mm) cryoprobe is compatible with the Ocutome-Fragmatome system. The iceball forms only at the tip and there is a rapid freeze and defrost cycle.
View Article and Find Full Text PDFFour patients developed progressive corneal opacification five to eight months after ultrasonic cataract extraction. Transmission electron microscopy was performed in all patients. Scanning electron microscopy, performed in only one patient, showed radiating folds of Descemet's membrane with pits on its posterior surface as well as focal traumatic disruption probably caused by probe contact.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!