A capillary electrophoresis-mass spectrometry method was used to analyze naphthenic acids in produced water samples. It was possible to detect cyclopentanecarboxylic, benzoic, cyclohexanebutyric, 1-naphthoic, decanoic, 3,5-dimethyladamantane-1-carboxylic, 9-anthracenecarboxylic, and pentadecanoic acids within ca. 13 min using a buffer composed of 40 mmol/L ammonium hydroxide, 32 mmol/L acetic acid and 20% v/v isopropyl alcohol, pH 8.6. The proposed method showed good repeatability, with relative standard deviation (RSD) values of 6.6% for the sum of the peak areas and less than 2% for the analysis time. In the interday analysis, the RSD values for the sum of the peak areas and migration time were 10.3% and 10%, respectively. The developed method demonstrated linear behavior in the concentration range between 5 and 50 mg/L for benzoic, decanoic, 3,5-dimethyladamantane-1-carboxylic and 9-anthracenecarboxylic acids, and between 10 and 50 mg/L for cyclopentanecarboxylic, cyclohexanebutyric, 1- naphthoic, and pentadecanoic acids. The detection limits values ranged from 0.31 to 1.64 mg/L. Six produced water samples were analyzed and it was possible to identify and quantify cyclopentanecarboxylic, benzoic, cyclohexanebutyric, and decanoic acids. The concentrations varied between 4.8 and 98.9 mg/L, proving effective in the application of complex samples.