Improving Quantitative Accuracy in Nontargeted Lipidomics by Evaluating Adduct Formation.

Anal Chem

West Coast Metabolomics Center, 451 Health Sci. Drive, University of California Davis, Davis, California 95616, United States.

Published: August 2023

AI Article Synopsis

  • Accurate and reproducible quantification of endogenous lipids is essential for lipidomic studies, but adduct formation during mass spectrometry can complicate this process.
  • Different types of diacylglycerol (DAG) adduct ions vary in abundance and ratio across various biological matrices, leading to potential inaccuracies in lipid quantification.
  • Combining multiple adduct ion types (specifically [M + NH] and [M + Na]) significantly improves quantification accuracy, reducing estimation errors compared to relying on a single adduct ion.

Article Abstract

For large-scale lipidomic analyses, accurate and reproducible quantification of endogenous lipids is crucial for comparing results within and across studies. Many lipids present in liquid chromatography-electrospray ionization-mass spectrometry form various adducts with buffer components. The mechanisms and conditions that dictate adduct formation are still poorly understood. In a positive mode, neutral lipids like mono-, di-, and triacylglycerides and cholesteryl esters typically generate [M + NH] adduct ions, although [M + Na], [M + K], and other (more complex) species can also be significantly abundant in MS1 precursor ion spectra. Variations in the ratios of these adducts (within and between matrices) can lead to dramatic inaccuracies during quantification. Here, we examine 48 unique diacylglycerol (DAG) species across 2366 mouse samples for eight matrix-specific data sets of plasma, liver, kidney, brain, heart muscle, gastrocnemius muscle, gonadal, and inguinal fat. Typically, no single adduct ion species accounted for more than 60% of the total observed abundance across each data set. Even within a single matrix, DAGs showed a high variability of adduct ratios. The ratio of [M + NH] adduct ions was increased for longer-chain DAGs and for polyunsaturated DAGs, at the expense of reduced ratios of [M + Na] adducts. When using three deuterated internal DAG standards, we found that absolute concentrations were estimated with up to 70% error when only one adduct ion was used instead of all adducts combined. Importantly, when combining [M + NH] and [M + Na] adduct ions, quantification results were within 5% accuracy compared to all adduct ions combined. Additional variance can be caused by other factors, such as instrument conditions or matrix effects.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11639661PMC
http://dx.doi.org/10.1021/acs.analchem.3c01221DOI Listing

Publication Analysis

Top Keywords

adduct ions
16
adduct
9
adduct formation
8
nh] adduct
8
adduct ion
8
improving quantitative
4
quantitative accuracy
4
accuracy nontargeted
4
nontargeted lipidomics
4
lipidomics evaluating
4