Objectives: The maximum output provided by a bone conduction (BC) device is one of the main factors that determines the success when treating patients with conductive or mixed hearing loss. Different approaches such as sound pressure measurements using a probe microphone in the external auditory canal or a surface microphone on the forehead have been previously introduced to determine the maximum output of active transcutaneous BC devices that are not directly accessible after implantation. Here, we introduce a method to determine the maximum output hearing level (MOHL) of a transcutaneous active BC device using patients' audiometric data.

Design: We determined the maximum output in terms of hearing level MOHL (dB HL) of the Bonebridge using the audiometric and direct BC threshold of the patient together with corresponding force levels at hearing threshold and the maximum force output of the device. Seventy-one patients implanted with the Bonebridge between 2011 and 2020 (average age 45 ± 19 years ranging from 5 to 84 years) were included in this study. The analyses of MOHLs were performed by (1) dividing patients into two groups with better or worse average audiometric BC threshold (0.5, 1, 2, 4 kHz), on the ipsilateral side or (2) by separating the MOHLs based on better or worse frequency-by-frequency specific audiometric BC thresholds on the ipsilateral (implanted) side.

Results: When using a frequency-by-frequency analysis obtained average ipsilateral MOHLs were in the range between 51 and 73 dB HL for frequencies from 0.5 to 6 kHz in the group with better audiometric BC threshold on the ipsilateral ears. The average contralateral MOHLs in the group with better contralateral hearing were in the range from 43 to 67 dB HL. The variability of the data was approximately 6 to 11 dB (SDs) across measured frequencies (0.5 to 6 kHz). The average MOHLs were 4 to 8 dB higher across frequencies in the group with better audiometric BC threshold on the ipsilateral ears than in the group with better audiometric BC threshold on the contralateral ears. The differences between groups were significant across measured frequencies ( t test; p < 0.05).

Conclusions: Our proposed method demonstrates that the individual frequency-specific MOHL on the ipsilateral and contralateral side of individual patients with a transcutaneous BC device can be determined mainly using direct and audiometric BC threshold data of the patients from clinical routine. The average MOHL of the implant was found 4 to 8 dB higher on the ipsilateral (implanted) side than on the contralateral side.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10718211PMC
http://dx.doi.org/10.1097/AUD.0000000000001415DOI Listing

Publication Analysis

Top Keywords

maximum output
20
audiometric threshold
20
group better
16
determine maximum
12
better audiometric
12
method determine
8
individual patients
8
active transcutaneous
8
bone conduction
8
clinical routine
8

Similar Publications

In the past few decades, any type of organization, from factories to government organizations, the banking sector, or educational institutions concentrates on increasing profit margins. To achieve this, one of the key factors is to achieve maximum output with minimum resources (input). Therefore, having an optimal plan to apply the resources has become extremely important for organizations.

View Article and Find Full Text PDF

The purpose was to assess whether visual feedback of torque contributes to motor unit (MU) firing rate reduction observed during post-activation potentiation (PAP) of skeletal muscle. From 15 participants 23 MUs were recorded with intramuscular fine-wire electrodes from the tibialis anterior during isometric dorsiflexion contractions at 20% of maximum, with and without both PAP and visual feedback of torque. A 5s maximal voluntary contraction (MVC) was used to induce PAP, and evoked twitch responses were assessed before and after.

View Article and Find Full Text PDF

Balancing pH and Pressure Allows Boosting Voltage and Power Density for a H-I Redox Flow Battery.

ACS Appl Energy Mater

January 2025

Department of Chemical Engineering, Delft University of Technology, Van der Maasweg 9, 2629HZ Delft, The Netherlands.

The decoupled power and energy output of a redox flow battery (RFB) offers a key advantage in long-duration energy storage, crucial for a successful energy transition. Iodide/iodine and hydrogen/water, owing to their fast reaction kinetics, benign nature, and high solubility, provide promising battery chemistry. However, H-I RFBs suffer from low open circuit potentials, iodine crossover, and their multiphase nature.

View Article and Find Full Text PDF

During the Fukushima Daiichi Nuclear Power Plant (FDNPP) accident on March 11, 2011, radionuclides such as tritium were released into the environment across Japan, obscuring the natural background signal of tritium in precipitation. This anthropogenic component was rapidly washed out by precipitation according to measurements in Japan. However, the impact of the accident on the natural tritium-based estimation of water system transit times in Fukushima and other prefectures in Japan remains uncertain.

View Article and Find Full Text PDF

1. Analysis of co-occurrence data with traditional indices has led to many problems such as sensitivity of the indices to prevalence and the same value representing either a strong positive or strong negative association across different datasets. In our recent study (Mainali et al 2022), we revealed the source of the problems that make the traditional indices fundamentally flawed and unreliable-namely that the indices in common use have no target of estimation quantifying degree of association in the non-null case-and we further developed a novel parameter of association, alpha, with complete formulation of the null distribution for estimating the mechanism of affinity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!