Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Rationale: Isolation of underivatized amino acids (AAs) using high-performance liquid chromatography (HPLC) is becoming a popular method for carbon (δ C) and nitrogen isotope (δ N) analyses of AAs because of the high analytical precision and for performing dual-isotope analysis. However, some AAs in natural samples, especially small, hydrophilic AAs, are not suitably separated using reversed-phase columns (e.g., C18) and ion-exchange columns (e.g., Primesep A).
Methods: We developed a new method for HPLC using a porous graphitic carbon column for the separation of nine hydrophilic AAs. After purification, δ C and δ N values of AAs were determined using elemental analyzer/isotope ratio mass spectrometry (EA/IRMS). We demonstrated the application of this method by determining δ C and δ N values of individual hydrophilic AAs in a biological sample, the muscle of blue mackerel (Scomber australasicus).
Results: Chromatographically, the baseline separation of hydrophilic AAs was achieved in both the standard mixture and the biological sample. We confirmed that δ C and δ N values of AA standards remained unchanged during the whole experimental procedure. The δ C values of AAs in mackerel muscle are also in good agreement with the values obtained using another verified method for δ C analysis.
Conclusions: The good separation performance of hydrophilic AAs and the reliability of δ C and δ N analyses of individual AAs using the porous graphite column offer a significant advantage over conventional settings. We suggest that, in the future, the HPLC × EA/IRMS method can be used for reliable δ C and δ N analyses of AAs in natural samples.