We conducted a comparative study of two models of point-of-care ultrasound devices for measuring post-void residual urine (PVRU). We prospectively enrolled 55 stroke inpatients who underwent both real-time B-mode ultrasound (Device A) and automated three-dimensional (3D) scanning ultrasound (Device B), with a total of 108 measurements. The median PVRU volume of Device B was 40 mL larger than that of Device A. The PVRU difference between the devices was positively and linearly correlated with PVRU. The correlation of PVRU volume between the devices was strong, but the agreement level was only moderate. Measurement deviations were observed in 43 (40%) and 11 (10%) measurements with Device B and Device A, respectively. The PVRU volume was low in spherical bladder shapes but sequentially increased in triangular, undefined, ellipsoid, and cuboid bladder shapes. Further comparison of 60 sets of PVRU without measurement deviations revealed higher agreements between the devices at correction coefficients of 0.52, 0.66, and 0.81 for PVRU volumes of <100, 100-200, and >200 mL, respectively. The automated 3D scanning ultrasound is more convenient for learning and scanning, but it exhibits larger measurement deviations. Real-time B-mode ultrasound accurately visualizes the urinary bladder but tends to underestimate the urinary bladder when the PVRU volume is large. Hence, real-time B-mode ultrasound with automated PVRU-based adjustment of calculation formulas may be a better solution for estimating bladder volume.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10417127 | PMC |
http://dx.doi.org/10.3390/diagnostics13152599 | DOI Listing |
Arab J Urol
July 2024
Faculty of Medicine, Urology Department, Ain Shams University, Cairo, Egypt.
Introduction And Objectives: Laser vaporization techniques have emerged as a prominent alternative to transurethral prostate resection in managing benign prostatic obstruction (BPO). This study focuses on assessing the effectiveness of the ejaculatory preserving laser vaporization of the prostate technique compared to the conventional non-ejaculatory approach in managing BPO.
Patients And Methods: Our study was performed between August 2022 and September 2023.
Prostate Int
March 2024
Department of Urology, Ng Teng Fong General Hospital, National University Health System, Singapore.
Background: Endoscopic enucleation of the prostate (EEP) has gained acceptance as an equitable alternative to transurethral resection of the prostate for benign prostate hyperplasia (BPH). Our primary aim is to compare peri-operative outcomes of EEP using thulium fiber laser (TFL) against high-power holmium laser (HPHL) in hands of experienced surgeons for large prostates (≥80 ml in volume). Secondary outcomes were assess complications within 1 year of follow up.
View Article and Find Full Text PDFDiagnostics (Basel)
August 2023
Stroke Center and Department of Neurology, Taipei Tzu Chi Hospital, Buddhist Tzu Chi Medical Foundation, New Taipei City 23142, Taiwan.
We conducted a comparative study of two models of point-of-care ultrasound devices for measuring post-void residual urine (PVRU). We prospectively enrolled 55 stroke inpatients who underwent both real-time B-mode ultrasound (Device A) and automated three-dimensional (3D) scanning ultrasound (Device B), with a total of 108 measurements. The median PVRU volume of Device B was 40 mL larger than that of Device A.
View Article and Find Full Text PDFMed J Malaysia
November 2022
Changi General Hospital, Department of Geriatric Medicine, Singapore.
Introduction: There is no consensus on the optimal postvoid residual urine volume (PVRU) as a cut-off value prior to performing intermittent catheterisation (IMC). We did a quality improvement project to determine a reasonable PVRU for use in the hospital setting.
Materials And Methods: All patients admitted to the five geriatric medicine wards in a geriatric department over a 5- month period who developed acute retention of urine were included in the project.
Int Braz J Urol
April 2021
Department of Urology, Hospital Universitario de Fuenlabrada, Madrid, Spain.
Objective: To generate high-quality data comparing the clinical efficacy and safety profile between monopolar transurethral resection of the prostate (M-TURP) and bipolar plasmakinetic resection of the prostate (PK-TURP) for benign prostatic hyperplasia (BPH).
Materials And Methods: Prospective, randomized, single-blinded study conducted in a tertiary-care public institution (Dec/2014-Aug/2016).
Inclusion Criteria: prostate of <80g in patients with drug-refractory lower urinary tract symptoms (LUTS), complications derived from BPH, or both.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!