Introduction: A fenestrated endograft (FE) is the first-line endovascular option for juxta and pararenal abdominal aortic aneurysms. A physician-modified stent-graft (PMSG) and laser in situ fenestration (LISF) have emerged to circumvent manufacturing delays, anatomic standards, and the procedure's cost raised by FE. The objective was to compare different fenestrations from a mechanical point of view.
Methods: In total, five Zenith Cook fenestrations (Cook Medical, Bloomington, IN, USA) and five Anaconda fenestrations (Terumo Company, Inchinnan, Scotland, UK) were included in this study. Laser ISF and PMSG were created on a Cook TX2 polyethylene terephthalate (PET) cover material (Cook Medical, Bloomington, IN, USA). In total, five LISFs and fifty-five PMSG were created. All fenestrations included reached an 8 mm diameter. Radial extension tests were then performed to identify differences in the mechanical behavior between the fenestration designs. The branch pull-out force was measured to test the stability of assembling with a calibrated 8 mm branch. Fatigue tests were performed on the devices to assess the long-term outcomes of the endograft with an oversized 9 mm branch.
Results: The results revealed that at over 2 mm of oversizing, the highest average radial strength was 33.4 ± 6.9 N for the Zenith Cook fenestration. The radial strength was higher with the custom-made fenestrations, including both Zenith Cook and Anaconda fenestrations (9.5 ± 4.7 N and 4.49 ± 0.28 N). The comparison between LISF and double loop PMSG highlighted a higher strength value compared with LISF (3.96 N ± 1.86 vs. 2.7 N ± 0.82; = 0.018). The diameter of the fenestrations varied between 8 and 9 mm. As the pin caliber inserted in the fenestration was 9 mm, one could consider that all fenestrations underwent an "elastic recoil" after cycling. The largest elastic recoil was observed in the non-reinforced/OC fenestrations (40%). A 10% elastic recoil was observed with LISF.
Conclusion: In terms of mechanical behavior, the custom-made fenestration produced the highest results in terms of radial and branch pull-out strength. Both PMSG and LISF could be improved with the standardization of the fenestration creation protocol.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10420147 | PMC |
http://dx.doi.org/10.3390/jcm12154911 | DOI Listing |
Eur J Vasc Endovasc Surg
January 2025
Aortic Center, Hôpital Marie Lannelongue, GHPSJ, Le Plessis-Robinson, France. Electronic address:
Objective: The aim of this study was to perform a comparative pre-clinical evaluation of a new fenestrated endovascular aneurysm repair dedicated stent using perfused 3D printed patient anatomies.
Methods: The test setup included a pulsatile pump set to reproduce human haemodynamics, four 3D printed pararenal aneurysms connected to a bench test, and four corresponding fenestrated grafts (Zenith Fenestrated; Cook, Bloomington, IN, USA). Bridging stents were sized based on analysis of patients computed tomography scans and included either four standard covered stents (BeGraft [BG]; Bentley, Hechingen, Germany) or four similar covered stent grafts mounted on a modified balloon to allow stent implantation and flaring in one step (BeFlared [BF]; Bentley).
Ann Thorac Surg Short Rep
December 2024
Division of Cardiac Surgery, Inova Heart and Vascular Institute, Inova Health Systems, Falls Church, Virginia.
Background: DeBakey type I aortic dissections (AD) are most frequently treated with hemiarch repair. A subset of patients demonstrates persistent distal end-organ ischemia secondary to persistent true lumen (TL) compression. We describe the use of bare metal stent grafting across the residual arch dissection with the Zenith Dissection Endovascular Stent (ZDES, Cook Medical) in 7 patients with type I AD that was repaired in a hemiarch configuration with a compromised distal TL and organ malperfusion.
View Article and Find Full Text PDFExpert Rev Med Devices
December 2024
Operative Unit of Vascular Surgery, IRCCS Policlinico San Donato, San Donato Milanese, Italy.
Introduction: Endovascular techniques and materials have significantly expanded their application in the treatment of abdominal and thoraco-abdominal aortic lesions, allowing for the management of increasingly complex pathologies that may require cannulation of target vessels. The treatment of such diseases deserves a particular approach and dedicated materials, for which correct procedural planning is mandatory. In the last decades, the use of 3D printing technology as an assisting tool for preoperative rehearsal of complex cases has progressively widespread.
View Article and Find Full Text PDFJ Endovasc Ther
November 2024
Department of Vascular and Endovascular Surgery, University Hospital Heidelberg, Heidelberg, Germany.
Purpose: This study investigates the accuracy of fully-automated maximum aortic diameter measurements in abdominal aortic aneurysm (AAA) patients using artificial intelligence software (PRAEVAorta 2, Nurea, Bordeaux, France).
Materials And Methods: This is a multicenter, retrospective validation study using prospectively collected data from the Zenith alpha for aneurysm Repair Registry (ZEPHYR). Automated measurements of PRAEVAorta 2 are compared with measurements of an internationally recognized core laboratory (Syntactx, New York, New York State).
EJVES Vasc Forum
September 2024
GEPROMED, Strasbourg, France.
Objective: The physical impact of the application of Heli-FX EndoAnchors (EA; Medtronic, Minneapolis, USA) on endograft (EG) material is unclear. This study aimed to examine the possible EG membrane alterations after EA implantation.
Methods: Heli-FX EndoAnchors were applied into four aortic endocuffs: AFX2 (Endologix Inc.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!