Background: This study compared the performance of near-infrared light transillumination (NILT; DIAGNOcam [DC]) in detecting occlusal caries lesions in deciduous molars with the performance of the International Caries Detection and Assessment System (ICDAS), digital radiographic method (RX) and laser fluorescence (DIAGNOdent pen [LFpen]).
Methods: Fifty-seven extracted deciduous molars with no frank cavitation caries lesions were selected. One site per tooth was evaluated twice each by two examiners using all methods. Then, a histological examination was performed using a stereomicroscope. The performance of the methods was assessed with respect to two thresholds: D (all visible caries lesions affecting enamel) and D (caries lesions extended into dentin).
Results: A weighted kappa and intraclass correlation coefficients were calculated for intra- and inter-observer reproducibilities, varying from 0.31 (RX) to 0.89 (LFpen). The intra-observer reproducibility at D and D thresholds varied from 0.51 (ICDAS) to 0.85 (DC) and from 0.30 (ICDAS) to 1.00 (RX), respectively. The inter-observer reproducibility varied from 0.28 (RX) to 0.47 (DC) and from 0.22 (LFpen) to 0.75 (DC), respectively. At the D threshold, higher sensitivity and accuracy were found for LFpen (0.76/0.71) and ICDAS (0.65/0.66); and higher specificities were found for RX (0.94), DC (0.79), and ICDAS (0.71). At the D threshold, all methods exhibited lower sensitivities (0.13-0.35) and higher specificity, accuracy, and area under the ROC curve values (0.70-1.00). All methods showed a correlation with histological results, especially for ICDAS and DC.
Conclusions: NILT technology (DC) and LFpen exhibited similar performance for the detection of occlusal caries lesions in deciduous molars when compared with ICDAS criteria.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.pdpdt.2023.103744 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!