Purpose: This prospective study evaluated the agreement among four optical coherence tomography angiography (OCTA) devices in the assessment of radial peripapillary capillary (RPC) density.
Methods: The study included 48 eyes of 48 subjects (14 healthy, 19 glaucomatous, and 15 non-glaucomatous optic neuropathy). Each participant was scanned using four OCTA devices in a random sequence: RTVue XR Avanti (RTVue), DRI OCT Triton (Triton), Revo NX 130 (Revo), and PLEX Elite 9000 (PlexE). All 6 × 6-mm grayscale OCTA images from each device were analyzed for RPC density using a customized algorithm. Agreement between each pair of devices was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman plots.
Results: There was a poor correlation between devices in all comparisons (RTVue-Triton, ICC = 0.34; RTVue-Revo, ICC = 0.31; RTVue-PlexE, ICC = 0.28; Triton-Revo, ICC = 0.31; Triton-PlexE, ICC = 0.17; Revo-PlexE, ICC = 0.34). Significant proportional biases (P < 0.05) and wide limits of agreement with apparent constant biases were identified in all comparisons. The mean difference was greatest for the RTVue-Revo pair (-49.3%; 95% confidence interval [CI], -52.9 to -45.8) and smallest for the Triton-PlexE pair (-7.7%; 95% CI, -10.1 to -5.3).
Conclusions: The RPC densities obtained from each device had poor inter-device agreement and significant biases and cannot be used interchangeably.
Translational Relevance: RPC density obtained from different OCTA devices is not interchangeable; thus, the progression of optic neuropathy should be monitored using the same OCTA device.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10424153 | PMC |
http://dx.doi.org/10.1167/tvst.12.8.7 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!