Introduction: A reduction in clot strength is a hallmark feature of trauma-induced coagulopathy. A better understanding of clot integrity can optimize resuscitation strategies. We designed a device to gauge clot strength by pressurizing fluids over a formed clot and measuring the pressure needed to dislodge the clot. We hypothesized that this device could distinguish between clots formed in hypocoagulable and hypercoagulable states by observing differences in the clot burst pressure.

Methods: Whole blood from healthy volunteers was collected into sodium citrate tubes and was treated with heparin or fibrinogen to generate clots in a hypocoagulable or hypercoagulable state, respectively. Small bore holes were drilled into polystyrene plates, and recalcified blood was pipetted into the holes. Plates were incubated at 37°C for 30 min to form clots. A pressure cap with an inlet for fluid from a syringe pump and an outlet leading to a measurement column was secured in the wells with a watertight seal.

Results: Clot burst pressure was normalized to individual baseline values to account for inherent differences in clot strength. The 1.0 g/L and 2.0 g/L fibrinogen groups were 1.65 ± 0.07 (P = 0.0078) and 2.26 ± 0.16 (P = 0.0078) times as strong as baseline, respectively. The 0.10, 0.15, or 0.20 USP units/mL groups were 0.388 ± 0.07 (P = 0.125), 0.31 ± 0.07 (P = 0.125), 0.21 ± 0.07 (P = 0.125) times as strong as baseline, respectively. Data were analyzed using Wilcoxon matched pairs signed rank testing.

Conclusions: This device tests clot strength using burst pressure, an easily interpreted clinical parameter not measured in existing devices. Future work can test blood from trauma patients to better understand trauma pathophysiology.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10626576PMC
http://dx.doi.org/10.1016/j.jss.2023.07.016DOI Listing

Publication Analysis

Top Keywords

clot strength
16
clot burst
12
burst pressure
12
clot
10
resuscitation strategies
8
hypocoagulable hypercoagulable
8
differences clot
8
times strong
8
strong baseline
8
pressure
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!