Coplanarity Constrained Ultrasound Probe Calibration Based on N-Wire Phantom.

Ultrasound Med Biol

School of Mathematical Sciences, Zhejiang University, Hangzhou, China. Electronic address:

Published: October 2023

AI Article Synopsis

Article Abstract

Objective: N-wire phantom-based ultrasound probe calibration has been used widely in many freehand tracked ultrasound imaging systems. The calibration matrix is obtained by registering the coplanar point cloud in ultrasound space and non-coplanar point cloud in tracking sensor space based on the least squares method. This method is sensitive to outliers and loses the coplanar information of the fiducial points. In this article, we describe a coplanarity-constrained calibration algorithm focusing on these issues.

Methods: We verified that the out-of-plane error along the oblique wire in the N-wire phantom followed a normal distribution and used it to remove the experimental outliers and fit the plane with the Levenberg-Marquardt algorithm. Then, we projected the points to the plane along the oblique wire. Coplanarity-constrained point cloud registration was used to calculate the transformation matrix.

Results: Compared with the other two commonly used methods, our method had the best calibration precision and achieved 25% and 36% improvement of the mean calibration accuracy than the closed-form solution and in-plane error method respectively at depth 16. Experiments at different depths revealed that our algorithm had better performance in our setup.

Conclusion: Our proposed coplanarity-constrained calibration algorithm achieved significant improvement in both precision and accuracy compared with existing algorithms with the same N-wire phantom. It is expected that calibration accuracy will improve when the algorithm is applied to all other N-wire phantom-based calibration procedures.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ultrasmedbio.2023.05.015DOI Listing

Publication Analysis

Top Keywords

n-wire phantom
12
point cloud
12
calibration
9
ultrasound probe
8
probe calibration
8
n-wire phantom-based
8
coplanarity-constrained calibration
8
calibration algorithm
8
oblique wire
8
calibration accuracy
8

Similar Publications

Coplanarity Constrained Ultrasound Probe Calibration Based on N-Wire Phantom.

Ultrasound Med Biol

October 2023

School of Mathematical Sciences, Zhejiang University, Hangzhou, China. Electronic address:

Objective: N-wire phantom-based ultrasound probe calibration has been used widely in many freehand tracked ultrasound imaging systems. The calibration matrix is obtained by registering the coplanar point cloud in ultrasound space and non-coplanar point cloud in tracking sensor space based on the least squares method. This method is sensitive to outliers and loses the coplanar information of the fiducial points.

View Article and Find Full Text PDF

Automatic spatial calibration of freehand ultrasound probe with a multilayer N-wire phantom.

Ultrasonics

February 2023

Beijing Engineering Research Center of Mixed Reality and Advanced Display, School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China.

The classic N-wire phantom has been widely used in the calibration of freehand ultrasound probes. One of the main challenges of the phantom is accurately identifying N-fiducials in ultrasound images, especially with multiple N-wire structures. In this study, a method using a multilayer N-wire phantom for the automatic spatial calibration of ultrasound images is proposed.

View Article and Find Full Text PDF

Ultrasound (US) imaging system is widely used in robotic systems for precision positioning in clinical applications. The US calibration is critical to minimize the difference of spatial coordinates between instruments, for minimally invasive surgery (MIS) in navigation systems. In this study, we propose a dual robotic arm system that combines US imaging with one arm for path planning and monitoring and accurate positioning with the other arm for instrument placement via the preplanning procedures.

View Article and Find Full Text PDF

A closed-form differential formulation for ultrasound spatial calibration: single wall phantom.

Ultrasound Med Biol

April 2015

Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia, Canada; Department of Mechanical Engineering, University of British Columbia, Vancouver, British Columbia, Canada. Electronic address:

Calibration is essential in freehand 3-D ultrasound to find the spatial transformation from the image coordinates to the sensor coordinate system. Ease of use, simplicity, precision and accuracy are among the most important factors in ultrasound calibration, especially when aiming to make calibration more reliable for day-to-day clinical use. We introduce a new mathematical framework for the simple and popular single-wall calibration phantom with a plane equation pre-determination step and the use of differential measurements to obtain accurate measurements.

View Article and Find Full Text PDF

A closed-form differential formulation for ultrasound spatial calibration: multi-wedge phantom.

Ultrasound Med Biol

September 2014

Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, Canada; Department of Mechanical Engineering, University of British Columbia, Vancouver, Canada. Electronic address:

Calibration is essential in freehand 3-D ultrasound to find the spatial transformation from the image coordinates to the sensor coordinate system. Calibration accuracy has significant impact on image-guided interventions. We introduce a new mathematical framework that uses differential measurements to achieve high calibration accuracy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!