Characterization of mechanical and radiation isocenter on an MR-guided radiotherapy (MRgRT) Linac.

J Appl Clin Med Phys

Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida, USA.

Published: November 2023

Background And Purpose: In the emerging paradigm of stereotactic radiosurgery being proposed for MR-guided radiotherapy (MRgRT), assessment of mechanical geometric accuracy is critical for the implementation of stereotactic delivery. We benchmarked the mechanical accuracy of an MR Linac system that lacks an onboard detector/array. Our mechanical tests utilize a half beam block (HBB) geometry that takes advantage of the sensitivity of a partially occluded detector.

Materials And Methods: Mechanical tests benchmarked the couch, MLC, and gantry geometric accuracy for an MR-Linac system. An HBB technique was used to irradiate an ionization chamber profiler (ICP) array with partial occlusion of individual detectors for characterization of MLC skew, beam divergence displacement, and RT isocenter localization. The sensitivity of the partially occluded detector's ICP-X (detector width) and ICP-Y (detector length) was characterized by displacing the detector relative to radiation isocenter by 0.2 mm increments, introduced through couch motion. The accuracy of the HBB ICP technique was verified with a starshot using radiochromic film, and the reproducibility was verified on a conventional C-arm Linac and compared to Winston-Lutz.

Results: The sensitivity of the HBB technique as quantified through the dose difference normalized to open field as a function of displacement from RT isocenter was 6.4%/mm and 13.0%/mm for the ICP-X and ICP-Y orientation, respectively, due to the oblong detector orientation. Couch positional accuracy and sag was within ±0.1 mm. Maximum MLC positional displacement was 0.7 mm with mean MLC skew at 0.07°. The maximum beam divergence displacement was 0.03 mm. The gantry angle was within 0.1°. Independent verification of the RT isocenter localization procedure produced repeatable results.

Conclusion: This work serves for characterizing the mechanical and geometric radiation accuracy for the foundation of an MR-guided stereotactic radiosurgery program, as demonstrated with high sensitivity and independent validation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10647948PMC
http://dx.doi.org/10.1002/acm2.14111DOI Listing

Publication Analysis

Top Keywords

radiation isocenter
8
mr-guided radiotherapy
8
radiotherapy mrgrt
8
stereotactic radiosurgery
8
mechanical geometric
8
geometric accuracy
8
mechanical tests
8
sensitivity partially
8
partially occluded
8
hbb technique
8

Similar Publications

Purpose: To evaluate the impact of patient setup errors on the dosimetry and radiobiological models of intensity-modulated radiotherapy (IMRT) for esophageal cancer.

Methods And Materials: This retrospective study with 56 patients in thermoplastic mask (TM) and vacuum bag (VB) groups utilized real setup-error (RSE) data from cone-beam CT scans to generate simulated setup-error (SSE) data following a normal distribution. The SSE data were applied to simulate all treatment fractions per patient by shifting the plan isocenter and recalculating the dose.

View Article and Find Full Text PDF

Purpose: To determine the dosimetric effects of set-up errors on boost coverage, and compares skin toxicity of sequential and simultaneous boost techniques for left-sided breast cancer.

Materials And Methods: This retrospective study included 23 early-stage breast cancer cases. Single isocenter HFWBI-SIB(s-SIB), single isocenter HFWBI-SB(s-SB) and dual isocenter HFWBI-SB(d-SB) were planing.

View Article and Find Full Text PDF

AI-based automatic patient positioning in a digital-BGO PET/CT scanner: efficacy and impact.

EJNMMI Phys

January 2025

Department of Nuclear Medicine, Rambam Health Care Campus, P.O.B. 9602, 3109601, Haifa, Israel.

Background: A recently released digital solid-state positron emission tomography/x-ray CT (PET/CT) scanner with bismuth germanate (BGO) scintillators provides an artificial intelligence (AI) based system for automatic patient positioning. The efficacy of this digital-BGO system in patient placement at the isocenter and its impact on image quality and radiation exposure was evaluated.

Method: The digital-BGO PET/CT with AI-based auto-positioning was compared (χ, Mann-Whitney tests) to a solid-state lutetium-yttrium oxyorthosilicate (digital-LYSO) PET/CT with manual patient positioning (n = 432 and 343 studies each, respectively), with results split into groups before and after the date of a recalibration of the digital-BGO auto-positioning camera.

View Article and Find Full Text PDF

Introduction: SRS for the treatment of limited brain metastases (BM) is widely accepted, but there are still limitations in the management of numerous BM. Frameless single-isocenter multitarget SRS is a novel technique that allows for rapid treatment delivery to multiple BM. We report our preliminary clinical, dosimetric, and patient´s shifts outcomes with this technique.

View Article and Find Full Text PDF

Introduction The Synaptive magnetic resonance imaging (MRI) system (Synaptive Medical, Toronto, Canada) is a midfield 0.5 T head-only scanner for imaging the head and neck in adults and pediatrics. The system received US FDA and Health Canada clearance for clinical use in 2020.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!