Quantitative assessment of image artifacts from zygoma implants on CBCT scans using different exposure parameters.

Proc Inst Mech Eng H

Department of Oral and Maxillofacial Surgery, Ministry of Health Turkey, Uskudar, Istanbul, Turkey.

Published: September 2023

This study was aimed at quantifying artifacts from zygoma implants in cone-beam computed tomography (CBCT) images using different exposure parameters. Two cadaver heads, one with two zygoma implants on each side and the other for control, were scanned using 18 different exposure parameters. Quantitative analysis was performed to evaluate the hypodense and hyperdense artifact percentages calculated as the percentage of the area. Hyperdense artifacts and hypodense artifacts were detected, followed by the calculation of the hyperdense and hypodense artifact percentages in the image. In the qualitative analysis of the artifacts, the scores used were as follows: absence (0), moderate presence (1), or high presence (2) for hypodense halos, thin hypodense lines, and hyperdense lines. Artifact analysis was performed qualitatively and quantitatively using the post-hoc Tukey and Two-way ANOVA tests. As a result, in the qualitative analyses, zygoma implants showed a significant difference compared to the control group with regard to hyperdense and hypodense artifacts ( < 0.05). There was a significant difference between the means according to the FOV size arithmetic averages ( < 0.05). In terms of voxel size, the difference was found to be significant, where 400 microns showed the highest hypodense artifact while 200 microns showed the lowest hypodense artifact. In conclusion, hypodense and hyperdense artifacts were significantly higher in cadavers with zygoma implants than in controls. As FOV and voxel size increase, more hypodense artifacts are produced by zygoma implants so smaller FOV and voxel sizes should be used to prevent poor image quality of adjacent teeth.

Download full-text PDF

Source
http://dx.doi.org/10.1177/09544119231190447DOI Listing

Publication Analysis

Top Keywords

zygoma implants
16
exposure parameters
12
artifacts zygoma
8
analysis performed
8
artifact percentages
8
hypodense artifacts
8
hyperdense hypodense
8
artifacts
6
hypodense
6
hyperdense
5

Similar Publications

Real-time surgical navigation systems (dynamic computer-aided surgery, d-CAIS) and static guided surgery (static computer-aided surgery, s-CAIS) have been shown to enhance the accuracy of zygomatic implant (ZI) placement. The objective of this systematic review was to evaluate and compare the accuracy and risk of complications associated with d-CAIS and s-CAIS in ZI placement. A systematic review of published studies involving more than 4 patients was conducted to assess and compare the accuracy of d-CAIS and s-CAIS in zygomatic implant placement.

View Article and Find Full Text PDF
Article Synopsis
  • The study focuses on infrazygomatic miniscrew implants (IZC-MSI) as temporary anchorage devices in orthodontics, highlighting their importance in supporting maxillary tooth movement.
  • The systematic review aimed to evaluate the success rate and stability factors of IZC-MSI through a comprehensive electronic search of multiple databases.
  • Results from 14 studies indicated a 92% overall success rate for these implants, with variations based on sample size and significant heterogeneity in factors influencing their stability.
View Article and Find Full Text PDF

Purpose: Placement of zygomatic implants in the most optimal prosthetic position is considered challenging due to limited bone mass of the zygoma, limited visibility, length of the drilling path and proximity to critical anatomical structures. Augmented reality (AR) navigation can eliminate some of the disadvantages of surgical guides and conventional surgical navigation, while potentially improving accuracy. In this human cadaver study, we evaluated a developed AR navigation approach for placement of zygomatic implants after total maxillectomy.

View Article and Find Full Text PDF

This study aimed to compare the angular, platform, and apical deviation of zygoma implants placed with the aid of a dynamic navigation (DN) device compared with the implants placed freehand. Eight cadaver heads were used for the study. Preoperative cone beam computerized tomography (CBCT) scans were made for the heads, and an implant-planning software was used to plan zygoma implants bilaterally in each scanned head.

View Article and Find Full Text PDF

Anatomical Measurements of the Malar Bone for Safe Zygomatic Implant Placement: A Study on Donated Bodies.

J Clin Med

November 2024

Cellular Signaling Laboratory, Anatomy Center, Department of Biomedical and Neuromotor Sciences (DIBINEM), University of Bologna, 40126 Bologna, Italy.

: The malar bone provides an anchorage point for zygomatic implants, avoiding invasive reconstructive surgeries in the fixed rehabilitation of fully edentulous and severely atrophic maxillae. The limited bone volume, however, requires precise implant placement to prevent complications related to nearby anatomical structures. This observational cross-sectional study aims to measure the malar and zygomatic arch bones and their distances from critical anatomical landmarks to guide surgeons in safe zygomatic implant placement.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!