Aims: Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty.
Methods: Patients recorded by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) between 1 September 1999 and 31 December 2020 who underwent hemiarthroplasty for the treatment of a hip fracture with the following implants were included: a cemented monoblock Exeter Trauma Stem (ETS), cemented Exeter V40 with a bipolar head, a monoblock Thompsons prosthesis (Cobalt/Chromium or Titanium), and an Exeter V40 with a Unitrax head. Overall and age-defined cumulative revision rates were compared over the ten years following surgery.
Results: A total of 41,949 hemiarthroplasties were included. Exeter V40 with a Unitrax head was the most commonly used (n = 20,707, 49.4%). The overall rate of revision was small. A total of 28,201 patients (67.2%) were aged > 80 years. There were no significant differences in revision rates across all designs of hemiarthroplasty in patients of this age at any time. The revision rates for all designs were < 3.5%, three years postoperatively. At subsequent times the ETS and Exeter V40 with a bipolar head performed well in all age groups. The unadjusted ten-year mortality rate for the whole cohort was 82.2%.
Conclusion: There was no difference in implant survival between all the designs of hemiarthroplasty in the first three years following surgery, supporting the selection of a cost-effective design of hemiarthroplasty for most patients with an intracapsular fracture of the hip, as determined by local availability and costs. Beyond this, the ETS and Exeter bipolar designs performed well in all age groups.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1302/0301-620X.105B8.BJJ-2022-1150.R3 | DOI Listing |
ANZ J Surg
September 2024
Department of Orthopaedic and Trauma Surgery, St George Hospital, Sydney, New South Wales, Australia.
Background: Early revision for total hip arthroplasty is a serious adverse outcome. There are multiple contributing risk factors for early revision. Risk factors can exist at the level of the surgeon and the level of the institution.
View Article and Find Full Text PDFArch Orthop Trauma Surg
May 2024
Department of Orthopaedic Surgery, Tauranga Hospital, Bay of Plenty, Private Bag 12024, Tauranga, 3143, New Zealand.
Background: Routine total hip arthroplasty (THA) using a short cemented stem as compared with a standard length cemented stem may have benefits in terms of stress distribution, bone preservation, stem subsidence and ease of revision surgery. Two senior arthroplasty surgeons transitioned their routine femoral implant from a standard 150 mm Exeter V40 cemented stem to a short 125 mm Exeter V40 cemented stem for all patients over the course of several years. We analysed revision rates, adjusted survival, and PROMS scores for patients who received a standard stem and a short stem in routine THA.
View Article and Find Full Text PDFBone Joint J
February 2024
Department of Orthopaedics, Hässleholm Hospital, Hässleholm, Sweden.
Aims: The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface.
Methods: Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem.
The aim is to review clinical and radiological outcomes for all cases of primary and revision THA, combining a cemented stem (Exeter V40) with a dual mobility component from a different manufacturer (SERF Novae), to evaluate whether concerns regarding mixing components from different manufacturers are justified. We identified 72 hip replacements performed between May 2010 and December 2015 using the SERF Novae dual mobility cup with an Exeter V40 stem, the majority of which were cemented (90%) and revisions (58%). Patients were evaluated clinically and radiologically at a minimum of two years.
View Article and Find Full Text PDFInt Orthop
March 2024
Department of Orthopaedics, Saint George University Teaching and County Hospital Centre for Musculoskeletal Surgery, Székesfehérvár, Hungary.
Purpose: A total of 894 hips were evaluated to describe the survivorship of Exeter cemented femoral stems and report the outcomes and complications of our 'Exeter- era', and there is no study from Central or Eastern Europe demonstrating similar results.
Methods: Between January 2000 and December 2009, a total of 894 hips were included who underwent Exeter universal and V40 femoral stems with a mean follow up of 13 years. Cemented Exeter low profile polyethylene cups were used in 889 patients (99.