This article reports a rare case of a total knee arthroplasty femoral component fracture. Fractures of early knee systems were attributed to design flaws. Modern design failures have been attributed to poor surgical technique or underlying osteolysis. Here, we report a fracture in the Vanguard prosthesis (Zimmer Biomet, Warsaw, IN) 12 years after implantation in a patient with a persistent flexion contracture. The fracture likely occurred due to fatigue failure of the anterior flange secondary to increased stress from a high riding patella. Although femoral component fractures are rare, they should be considered as a potential complication, especially in patients with special load considerations. For these patients, it is essential that the prosthesis be properly supported with clean cuts and an adequate cement mantle.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374856 | PMC |
http://dx.doi.org/10.1016/j.artd.2023.101174 | DOI Listing |
J Arthroplasty
January 2025
Georgetown University School of Medicine, MedStar Georgetown University Hospital, Washington, D.C, USA. Electronic address:
Introduction: As the number of revision total knee arthroplasties (rTKA) continues to rise, there is increasing interest in the use of contemporary rotating hinge prostheses. These devices often incorporate porous cones to fill bone defects and enhance long-term fixation. This study evaluated the clinical and functional outcomes and survivorship in rTKA patients utilizing a rotating hinge prosthesis with flexible titanium (FT) cones, porous tantalum (PT) cones, or no cones.
View Article and Find Full Text PDFPharmaceuticals (Basel)
January 2025
Smart Farm Research Center, Korean Institute of Science and Technology (KIST), Gangneung 25451, Republic of Korea.
Background: Osteoporosis is characterized by the microstructural depletion of bone tissue and decreased bone density, leading to an increased risk of fractures. Nakai, an endemic species of the Korean Peninsula, grows wild in Ulleungdo. In this study, we aimed to investigate the effects of and its components on osteoporosis.
View Article and Find Full Text PDFJ Clin Med
January 2025
Department of Mechanical Engineering, Yonsei University, Seoul 03722, Republic of Korea.
: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) uses preoperative three-dimensional imaging to create cutting blocks tailored to patient anatomy. However, there is debate regarding the additional benefits of PSI in terms of improved alignment or functional outcomes compared to using conventional instruments. Although PSI design has undergone continuous development, the improvements have not been incorporated.
View Article and Find Full Text PDFMedicina (Kaunas)
January 2025
Department of Orthopedic Surgery, Keimyung University Dongsan Hospital, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.
A basicervical femoral fracture is a relatively uncommon type of proximal femoral fracture. However, as the proportion of proximal femoral fractures rises in conjunction with the aging of society, the absolute number of patients with basicervical femoral fractures is also increasing. Nevertheless, the optimal surgical methods for the treatment of basicervical femoral fractures remain a topic of debate.
View Article and Find Full Text PDFFibrous dysplasia (FD) is a benign tumor condition in which normal bone is replaced by structurally deficient fibrous lamellar bone. It represents approximately 5-7% of benign bone tumors and occurs in two presentations: monostotic, which is the most common, and polyostotic. The proximal femur is one of the most common locations for benign tumors, including FD.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!