Background: Electrical lead abnormalities (ELAs) can result in device malfunction, leading to significant morbidity in patients with cardiac implantable electronic devices (CIEDs).

Objective: We sought to determine the prevalence and management of ELAs in patients with CIEDs.

Methods: This was a retrospective cohort study of patients implanted with a CIED between 2012 and 2019 at a tertiary care center. The primary outcome was ELA defined as increased capture threshold (≥2× implantation value), decreased sensing (≤0.5 implantation value), change in impedance (>50% over 3 months), or nonphysiologic potentials. A secondary outcome of device clinic utilization was also collected.

Results: There were 2996 unique patients (35% female) included with 4600 leads (57% Abbott, 43% Medtronic). ELAs were observed in 135 (3%) leads, including 124 (92%) Abbott and 10 (7%) Medtronic leads (hazard ratio 9.25, .001). Mean follow-up was 4.5 ± 2.2 years. ELAs were associated smaller lead French size, atrial location, and Abbott leads. Lead revision was required in 28% of cases. Patients with lead abnormalities had 38% more in-clinic visits per patient year of follow-up compared with those without ( .001).

Conclusion: ELAs were more frequent in certain models, which increased rates of revision and follow-up. Identification of factors that mitigate these abnormalities to improve lead performance are required to improve care for these devices and provide efficient healthcare.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10373148PMC
http://dx.doi.org/10.1016/j.hroo.2023.05.002DOI Listing

Publication Analysis

Top Keywords

lead abnormalities
12
prevalence management
8
electrical lead
8
cardiac implantable
8
implantable electronic
8
lead
6
leads
5
elas
5
patients
5
management electrical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!