Purpose: To study the feasibility of optimizing the Clinical Target Volume to Planning Target Volume (CTV-PTV) margin in prostate radiotherapy(RT) with a general-purpose linear accelerator using an in-house developed position monitoring system, SeedTracker.
Methods: A cohort of 30 patients having definitive prostate radiotherapy treated within an ethics-approved prospective trial was considered for this study. The intrafraction prostate motion and the position deviations were measured using SeedTracker system during each treatment fraction. Using this data the CTV-PTV margin required to cover 90% of the patients with a minimum of 95% of the prescription dose to CTV was calculated using van Herk's formula. The margin calculations were performed for treatment scenarios both with and without applying the position corrections for observed position deviations. The feasibility of margin reduction with real-time monitoring was studied by assessing the delivered dose that incorporates the actual target position during treatment delivery and comparing it with the planned dose. This assessment was performed for plans generated with reduced CTV-PTV margin in the range of 7mm-3mm.
Results: With real-time monitoring and position corrections applied the margin of 2.0mm, 2.1mm and 2.1mm in LR, AP and SI directions were required to meet the criteria of 90% population to receive 95% of the dose prescription to CTV. Without position corrections applied for observed position deviations a margin of 3.1mm, 4.0mm and 3.0mm was required in LR, AP and SI directions to meet the same criteria. A mean ± SD reduction of 0.5 ± 1.8% and 3 ± 7% of V60 for the rectum and bladder can be achieved for every 1mm reduction of PTV margin. With position corrections applied, the CTV D99 can be delivered within -0.2 ± 0.3 Gy of the planned dose for plans with a 3mm margin. Without applying corrections for position deviations the CTV D99 was reduced by a maximum of 1.1 ± 1.1 Gy for the 3mm margin plan and there was a statistically significant difference between planned and delivered dose for 3mm and 4mm margin plans.
Conclusion: This study demonstrates the feasibility of reducing the margin in prostate radiotherapy with SeedTracker system without compromising the dose delivery accuracy to CTV while reducing dose to critical structures.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10373585 | PMC |
http://dx.doi.org/10.3389/fonc.2023.1116999 | DOI Listing |
BMC Musculoskelet Disord
January 2025
Department of Orthopedics, The Affiliated Suzhou Hospital of Nanjing Medical University, Suzhou Municipal Hospital, Gusu School, Nanjing Medical University, Suzhou, 215000, China.
Background: To analyze the effects of the positioning of a bolt in the femoral neck system (FNS) on the short-term outcomes of middle-aged and young adults with displaced femoral neck fractures (FNFs).
Methods: This was a retrospective study involving 114 middle-aged and young adults with displaced FNFs who were surgically treated with internal fixation via the FNS in the Department of Orthopedics, Suzhou Municipal Hospital, from December 2019 to January 2023. The degree of deviation of the central axis of the femoral head and neck from the tip of the bolt (W), the tip‒apex distance (TAD) and the length of femoral neck shortening (LFNS) were measured on postoperative X-ray and computed tomography (CT) scan images.
Int Ophthalmol
January 2025
Department of Ophthalmology, The Second Hospital of Jilin University, Changchun, Jilin, China.
Purpose: To evaluate the efficacy and complications of simplified graded inferior oblique anterior transposition (IOAT) in treating at least 10 PD vertical deviation in the primary position and inferior oblique muscle overaction (IOOA).
Methods: This retrospective study reviewed the medical records of 65 patients treated with simplified graded IOAT procedures for both vertical deviation and IOOA. Patients were grouped according to vertical deviation in the primary position.
Eur Spine J
January 2025
Department of Spine Surgery, The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang, 050051, China.
Objective: Spinopelvic sagittal balance ensures efficient posture and minimizes energy expenditure by aligning the spine, pelvis, and lower extremities. Deviations can cause clinical issues like back pain and functional limitations. Key radiographic parameters, including pelvic tilt (PT), pelvic incidence (PI), sacral slope (SS), and lumbar lordosis (LL), are essential for evaluating spinal pathologies and planning surgeries.
View Article and Find Full Text PDFTraffic Inj Prev
January 2025
Department of Civil Engineering, Indian Institute of Technology, Indore, India.
Objective: The study aims to develop continuous trajectory profiles along curves with minimal error. It also focuses on formulating a percentage trajectory transection rate model as a function of geometric parameters (e.g.
View Article and Find Full Text PDFJ Dent
January 2025
Key Laboratory of Shaanxi Province for Craniofacial Precision Medicine Research, College of Stomatology, Xi'an Jiaotong University; Clinical Research Center of Shaanxi Province for Dental and Maxillofacial Disease, College of Stomatology, Xi'an Jiaotong University; Department of Implant Dentistry, College of Stomatology, Xi'an Jiaotong University. Electronic address:
Objective: The study aimed to evaluate the accuracy and safety of the semi-active robotic system for implant placement in atrophic posterior maxilla.
Methods: Patients underwent robot-assisted implant placement in atrophic posterior maxilla were identified and included. Cone-beam computed tomography (CBCT) was performed before surgery.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!