We present a phase-locked synthetic wavelength interferometer that enables a complete elimination of cyclic errors in absolute distance measurements. With this method, the phase difference between the reference and measurement paths is fed back into a phase lock-in system, which is then used to control the synthetic wavelength and set the phase difference to zero using an external cavity acousto-optic modulator. We validated the cyclic error removal of the proposed phase-locked method by comparing it with the conventional phase-measuring method of the synthetic wavelength interferometer. By analyzing the locked error signal, we achieved a precision of 0.6 mrad in phase without any observed cyclic errors.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10384882PMC
http://dx.doi.org/10.3390/s23146253DOI Listing

Publication Analysis

Top Keywords

synthetic wavelength
16
wavelength interferometer
12
phase-locked synthetic
8
absolute distance
8
cyclic error
8
cyclic errors
8
phase difference
8
wavelength
4
interferometer femtosecond
4
femtosecond laser
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!