(1) Purpose: to compare right ventricular (RV) functional parameters in children with surgically repaired congenital heart disease (CHD) using single/double breath hold (BH) and free-breathing (FB) real-time compressed sensing (CS) cine cardiac magnetic resonance (cMRI) with standard retrospective segmented multi breath hold (RMB) cine cMRI. (2) Methods: Twenty patients with CHD underwent BH and FB, as well as RMB cine cMRI, at 3T to obtain a stack of continuous axial images of the RV. Two radiologists independently performed qualitative analysis of the image quality (rated on a 5-point scale; 1 = non-diagnostic to 5 = excellent) and quantitative analysis of the RV volume measurements. (3) Results: The best image quality was provided by RMB (4.5; range 2-5) compared to BH (3.9; range 3-5; = 0.04) and FB (3.6; range 3-5; < 0.01). The RV functional parameters were comparable among BH, FB, and RMB with a difference of less than 5%. The scan times for BH (44 ± 38 s, < 0.01) and FB (24 ± 7 s, < 0.01) were significantly reduced compared to for RMB (261 ± 68 s). (4) Conclusions: CS-FB and CS-BH real-time cine cMRI in children with CHD provides diagnostic image quality with excellent accuracy for measuring RV function with a significantly reduced scan time compared to RMB.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10377861PMC
http://dx.doi.org/10.3390/diagnostics13142403DOI Listing

Publication Analysis

Top Keywords

cine cmri
12
image quality
12
compressed sensing
8
congenital heart
8
heart disease
8
functional parameters
8
breath hold
8
rmb cine
8
range 3-5
8
compared rmb
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!