Numerous cardiac output (CO) technologies were developed to replace the 'gold standard' pulmonary artery thermodilution due to its invasiveness and the risks associated with it. Minimally invasive lithium dilution (LiD) shows excellent agreement with thermodilution and can be used as a reference standard in animals. This study evaluated CO via noninvasive electrical cardiometry (EC) and acquired hemodynamic variables against CO measured using LiD in six healthy, anesthetized dogs administered different treatments (dobutamine, esmolol, phenylephrine, and high-dose isoflurane) impacting CO values. These treatments were chosen to cause drastic variations in CO, so that fair comparisons between EC and LiD across a wide range of CO values (low, intermediate, and high) could be made. Statistical analysis included linear regression, Bland-Altman plots, Lin's concordance correlation coefficient (ρ), and polar plots. Values of < 0.05 represented significance. Good agreement was observed between EC and LiD, but consistent underestimation was noted when the CO values were high. The good trending ability, ρ of 0.88, and low percentage error of ±31% signified EC's favorable performance. Other EC-acquired variables successfully tracked changes in CO measured using LiD. EC may be a pivotal hemodynamic tool for continuously monitoring circulatory changes, as well as guiding and treating cardiovascular anesthetic complications in clinical settings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10376001PMC
http://dx.doi.org/10.3390/ani13142362DOI Listing

Publication Analysis

Top Keywords

electrical cardiometry
8
cardiac output
8
hemodynamic variables
8
lithium dilution
8
anesthetized dogs
8
measured lid
8
lid
5
evaluation electrical
4
cardiometry measuring
4
measuring cardiac
4

Similar Publications

Cardiac output can be estimated non-invasively by electrical cardiometry with the ICON® monitor (Osypka Medical GmbH, Berlin, Germany). Conflicting results have been reported regarding the cardiac output measurement performance of electrical cardiometry. In this prospective method comparison study, we compared cardiac output measured using electrical cardiometry (EC-CO; test method) with cardiac output measured using intermittent pulmonary artery thermodilution (PATD-CO; reference method) in patients after coronary artery bypass graft (CABG) surgery.

View Article and Find Full Text PDF

Objective: Scanty data are available about neonatal systemic vascular resistances (SVR). We aim to provide reference values and nomograms for neonatal SVR.

Design: Multicenter, cross-sectional,descriptive study performed in France and Italy.

View Article and Find Full Text PDF

Aim: Acute circulatory failure is commonly encountered in critically ill patients, that requires fluid administration as the first line of treatment. However, only 50% of patients are fluid-responsive. Identification of fluid responders is essential to avoid the harmful effects of overzealous fluid therapy.

View Article and Find Full Text PDF

This study aimed to evaluate TFC by EC versus lung ultrasound (LUS) findings for diagnosing and follow-up of TTN in late preterm and term neonates. This prospective observational study was conducted on 80 neonates with gestational age ≥ 34 weeks. TTN group included 40 neonates diagnosed with TTN, and no lung disease (NLD) group included 40 neonates without respiratory distress.

View Article and Find Full Text PDF