Background:  The implantation of a spacer is a common practice after anterior diskectomy in cervical spine. Polyether ether ketone (PEEK) cages have replaced titanium implants due to their better radiologic visibility and appearance in postoperative magnetic resonance imaging (MRI) scans. However, PEEK showed apparently higher nonunion rates than titanium cages. The aim of the study was to evaluate the fusion behavior of plain PEEK cages in comparison to titanium-coated PEEK (TiPEEK) cages.

Method:  We randomized 104 patients with single-level cervical radiculopathy or mild myelopathy. They were divided into two groups of 52 patients each, receiving either a PEEK cage or the titanium-coated variant of the same cage type. The 1- and 2-year follow-ups were completed by 43 patients in the PEEK group and by 50 patients in the TiPEEK group. Fusion was determined by plain X-ray and lateral functional X-ray.

Results:  Two years after surgery, a complete fusion was observed in 37 patients of the PEEK group (86%). Six cases were considered as nonunions. In the TiPEEK group, we found 41 fusions (82%) and 9 nonunions at this time. The difference was not considered significant ( = 0.59). The clinical evaluation of the two groups showed no difference in the neurologic examination as well in the pain scores over the time period.

Conclusions:  Despite some assumptions about an advantage of TiPEEK over PEEK cages for fusion in cervical spine surgery, this prospective randomized controlled study did not find an accelerated or improved fusion using TiPEEK for anterior cervical diskectomy.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-0043-1770694DOI Listing

Publication Analysis

Top Keywords

peek cages
20
peek
10
titanium-coated peek
8
anterior cervical
8
randomized controlled
8
controlled study
8
cervical spine
8
patients peek
8
peek group
8
tipeek group
8

Similar Publications

Article Synopsis
  • Cage subsidence can negatively affect lumbar fusion procedures, with material selection (PEEK vs. 3D-Ti) influencing this issue; the study aims to compare their subsidence rates.
  • The systematic search reviewed 265 patients from three high-quality studies, focusing on cage subsidence and classified subsidence rates using a specific method.
  • Results indicated that 3D-Ti cages have a significantly lower rate of subsidence compared to PEEK cages, with less severe subsidence and better overall performance.
View Article and Find Full Text PDF

Background Context: Cage subsidence is a complication of interbody fusion associated with poor clinical outcomes. 3D-printed titanium interbody cages allow for the alteration of features such as stiffness and porosity. However, the influence of these features on subsidence and their biological effects on fusion have not been rigorously evaluated.

View Article and Find Full Text PDF

Background: Chronic low back pain secondary to degenerative disc disease is a significant public health issue worldwide, contributing to substantial health care burdens and patient disability. Anterior lumbar interbody fusion (ALIF) has emerged as a promising surgical solution, offering benefits such as disc height restoration, reduced neural compression, and improved spinal alignment. This study evaluates the efficacy of stand-alone ALIF using polyetheretherketone (PEEK) cages, structural femoral head allografts, and recombinant human bone morphogenetic protein-2 (rhBMP-2) in treating discogenic low back pain caused by degenerative disc disease.

View Article and Find Full Text PDF

Structural design and biomechanical analysis of a combined titanium and polyetheretherketone cage based on PE-PLIF fusion.

Med Biol Eng Comput

October 2024

Shanxi Bethune Hospital, Shanxi Academy of Medical Sciences, Tongji Shanxi Hospital, Third Hospital of Shanxi Medical University, Taiyuan, 030032, China.

Article Synopsis
  • The study explores the biomechanical performance of a newly designed spinal fusion cage compared to traditional titanium and PEEK cages, highlighting the drawbacks of stress shielding and poor bone ingrowth associated with the former materials.
  • Using finite element analysis, the new combined cage showed a 9.5% increase in strain capacity and reduced endplate and fusion stresses by 12% and 17%, respectively, during dynamic movements.
  • The findings indicate that the combined cage may improve fusion stability and reduce adverse effects tied to titanium cages, offering insights for better clinical cage designs in spinal fusion surgeries.
View Article and Find Full Text PDF

Objective: Biportal endoscopic transforaminal lumbar interbody fusion (BE-TLIF) is an emerging, minimally invasive technique performed under biportal endoscopic guidance. However, concerns regarding cage subsidence and sufficient fusion during BE-TLIF necessitate careful selection of an appropriate interbody cage to improve surgical outcomes. This study compared the fusion rate, subsidence, and other radiographic parameters according to the material and size of the cages used in BE-TLIF.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!