Construction of a radiation beam scanner and investigation of volume averaging correction factor effects on beam-profile.

Radiat Prot Dosimetry

Measurements and environmental monitoring, Radiation and Nuclear Safety Authority (STUK), Jokiniemenkuja 1, Vantaa FI-01370, Finland.

Published: August 2023

In radiation beam-profile measurements, an accurate positioning of the detector with high position resolution is essential. For this purpose, we built a scanning device capable of moving a detector in three dimensions (3D) using mainly parts from a commercial 3D-printer. The accuracy and repeatability of movement was tested with caliper, laser displacement sensor and repeated 60Co beam-profile measurements in a water phantom. The results from the caliper and the sensor showed position accuracy for the scanner to be better than $\pm$150 μm. The standard deviation of the error in position from laser sensor measurements was approximately 30 μm, and the beam profile scans showed a maximum deviation from the mean position of 50 μm. The effect of volume averaging correction factors on 60Co beam-profile was investigated with two different sized ionization chambers. The differences in the profiles were reduced significantly after applying the correction factors.

Download full-text PDF

Source
http://dx.doi.org/10.1093/rpd/ncad198DOI Listing

Publication Analysis

Top Keywords

volume averaging
8
averaging correction
8
beam-profile measurements
8
60co beam-profile
8
correction factors
8
construction radiation
4
radiation beam
4
beam scanner
4
scanner investigation
4
investigation volume
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!